Centre Hospitalier Universitaire Grenoble Alpes, Médecine Intensive Réanimation, INSERM, Université Grenoble-Alpes, Grenoble, France.
Clinical Investigation Centre 1429, APHP, Hôpital Raymond-Poincaré, Garches, France.
Respir Care. 2023 Apr;68(4):462-469. doi: 10.4187/respcare.10086.
Mechanical insufflation-exsufflation (MI-E) devices are used to improve airway clearance in individuals with acute respiratory failure. Some MI-E devices measure cough peak flow (CPF) during MI-E to optimize pressure adjustments. The aim was to compare CPF and effective cough volume (ECV: volume expired/coughed > 3 L/s) measurements between 4 MI-E devices under simulated conditions of stable versus collapsed airway.
Four MI-E devices were tested on the bench. Each device was connected via a standard circuit to a collapsible tube placed in an airtight chamber that was attached to a lung model with adjustable compliance and resistance. Pressure was measured upstream and downstream the collapsing tube; air flow was measured between the chamber and the lung model. Each device was tested in 2 conditions: collapse condition (0 cm HO) and no-collapse condition (-70 cm HO). For each condition, 6 combinations of inspiratory/expiratory pressures were applied. CPF was measured at the "mouth level" by the device built-in flow meter and at the "tracheal level" by a dedicated pneumotachograph. Comparisons were performed with non-parametric tests.
CPF values measured at the tracheal level and ECV values differed between devices for each inspiratory/expiratory pressure in the collapse and no-collapse conditions ( < .001). CPF values were significantly lower at the tracheal level in the collapse as compared with the no-collapse condition ( < .001 for each device), whereas they were higher at the mouth level ( < .05) for 3 of the 4 devices.
CPF values differed significantly across MI-E devices, highlighting limitation(s) of using only CPF values to determine cough effectiveness. In simulated of airway collapse, CPF increased at the mouth, whereas it decreased at the tracheal level.
机械通气-呼气(MI-E)设备用于改善急性呼吸衰竭患者的气道清除能力。一些 MI-E 设备在 MI-E 期间测量咳嗽峰流速(CPF),以优化压力调整。目的是在稳定与塌陷气道的模拟条件下比较 4 种 MI-E 设备的 CPF 和有效咳嗽量(ECV:呼出/咳出的体积> 3 L/s)测量值。
在台架上测试了 4 种 MI-E 设备。每个设备都通过标准电路连接到放置在密封室中的可折叠管,该管连接到具有可调节顺应性和阻力的肺模型。在折叠管的上游和下游测量压力;在腔室和肺模型之间测量气流。每个设备在 2 种条件下进行测试:塌陷条件(0 cm HO)和无塌陷条件(-70 cm HO)。对于每种条件,应用 6 种吸气/呼气压力组合。CPF 通过设备内置的流量计在“口腔水平”测量,通过专用的呼吸量计在“气管水平”测量。使用非参数检验进行比较。
在塌陷和无塌陷条件下,对于每个吸气/呼气压力,CPF 值在气管水平和 ECV 值在不同设备之间均存在差异(<0.001)。与无塌陷条件相比,塌陷条件下 CPF 值在气管水平显著降低(每个设备<0.001),而在口腔水平则升高(对于 4 个设备中的 3 个<0.05)。
CPF 值在 MI-E 设备之间存在显著差异,这凸显了仅使用 CPF 值来确定咳嗽效果的局限性。在气道塌陷模拟中,CPF 在口腔处增加,而在气管水平处减少。