Friedman W A, Curran M T
Department of Neurological Surgery, University of Florida College of Medicine, Gainesville.
Neurosurgery. 1987 Nov;21(5):755-8. doi: 10.1227/00006123-198711000-00033.
A new method for improving the accuracy of somatosensory evoked potential (SEP) monitoring is described. Two extremities are stimulated at 5 Hz, with an interextremity stimulus interval of 100 ms. Information from a "control" extremity, such as an arm during scoliosis operation, may allow more rapid and accurate interpretation of changes in SEPs during intraoperative monitoring.
本文描述了一种提高体感诱发电位(SEP)监测准确性的新方法。以5Hz的频率刺激双侧肢体,双侧肢体刺激间隔为100ms。来自“对照”肢体的信息,比如脊柱侧弯手术中上肢的信息,可能有助于在术中监测期间更快速准确地解读SEP的变化。