In Situ Measurement of Acoustic Attenuation for Focused Ultrasound Ablation Surgery Using a Boiling Bubble at the Focus.

作者信息

Zhang Tianfeng, Zhou Yufeng, Wang Zhibiao

机构信息

State Key Laboratory of Ultrasound in Medicine and Engineering, College of Biomedical Engineering, Chongqing Medical University, Chongqing, China; Chongqing Key Laboratory of Biomedical Engineering, Chongqing Medical University, Chongqing, China.

State Key Laboratory of Ultrasound in Medicine and Engineering, College of Biomedical Engineering, Chongqing Medical University, Chongqing, China; Chongqing Key Laboratory of Biomedical Engineering, Chongqing Medical University, Chongqing, China.

出版信息

Ultrasound Med Biol. 2023 Jul;49(7):1672-1678. doi: 10.1016/j.ultrasmedbio.2023.02.014. Epub 2023 Mar 31.

Abstract

OBJECTIVE

Acoustic attenuation in the propagation path of focused ultrasound ablation surgery determines the energy loss toward the focal region and is critical to the consequent treatment outcomes. In situ non-invasive, reliable, and accurate measurement is challenging for multi-layered heterogeneous tissues within the focusing angle.

METHODS

A novel measurement approach is proposed and its performance is evaluated using ex vivo porcine tenderloin and bovine heart. A big boiling bubble (i.e., larger than a few millimeters in size) was produced at the focus as a strong reflector inside the tissue, and the echo amplitudes were used to determine the acoustic attenuation. Two models, acoustic ray and energy loss, were developed to derive the equivalent acoustic attenuation coefficient for a focused beam.

RESULTS

The measured acoustic attenuation coefficients of ex vivo porcine tenderloin and bovine heart at 0.97 MHz and a thickness of 3 cm are 0.159 ± 0.002 and 0.250 ± 0.005 Np/cm, respectively, which are all within the scope of measured values in the literature. In addition, the echo amplitude is sensitive to the conditions of the propagation path, and the inverse acoustic attenuation coefficient of the silicone gel pad placed in front of the tissue sample was 0.807 ± 0.002 Np/cm, which is comparable to the measurement using the insertion substitution method, 0.766 ± 0.003 Np/cm.

CONCLUSION

Our proposed approach could determine the tissue acoustic attenuation for focused ultrasound ablation surgery reliably and accurately in situ. The easy operating protocol may allow clinical translation and adoption for improved safety and efficacy.