Scatter rejection by electronic collimation.

作者信息

Plenkovich D, Sorenson J A, Kruger R A

出版信息

Med Phys. 1986 Mar-Apr;13(2):158-63. doi: 10.1118/1.595940.

Abstract

An electronic scanning-slit technology is under development which involves replacing paired fore and aft slits for scatter rejection with only one fore slit. As the slit scans across the portion of the patient to be imaged, pulsed x-ray exposures produce images of the slit in successive positions, which are digitized and stored in computer memory. Software techniques are used for tracking the slit image, and discriminating against scatter and veiling glare lying outside the slit image. Such "electronic collimation" does not require synchronization between the slit-scanning and detector readout, which makes it much simpler than alternative methods and potentially adaptable to any digital fluorography system. The performance characteristics of a prototype unit are described and images of a low-contrast phantom are presented.

摘要

一种电子扫描狭缝技术正在研发中,该技术涉及用仅一个前狭缝取代用于散射抑制的前后配对狭缝。当狭缝扫过要成像的患者部位时,脉冲X射线曝光会在连续位置产生狭缝图像,这些图像被数字化并存储在计算机内存中。软件技术用于跟踪狭缝图像,并区分狭缝图像之外的散射和蒙罩眩光。这种“电子准直”不需要狭缝扫描和探测器读出之间的同步,这使其比其他方法简单得多,并且有可能适用于任何数字荧光成像系统。描述了一个原型装置的性能特征,并展示了低对比度体模的图像。