Department of Neurosurgery, Queen Elizabeth Hospital Birmingham, Birmingham, B15 2TH, UK.
Institute of Metabolism and Systems Research, University of Birmingham, Birmingham, UK.
Acta Neurochir (Wien). 2023 Jun;165(6):1523-1531. doi: 10.1007/s00701-023-05572-z. Epub 2023 Apr 18.
The M.scio telesensor (Aesculap-Miethke, Germany) is a device integrated within a ventriculoperitoneal (VP) shunt for non-invasive measurement of the intracranial pressure (ICP). The purpose of this study was to analyze the telemetric recordings with the M.scio system in shunted patients with idiopathic intracranial hypertension (IIH), in order to determine reference values and assist the interpretation of telemetric data.
This was a cohort study of consecutive patients with fulminant IIH who underwent primary VP shunt insertion between July 2019 and June 2022. The first telemetric measurements after surgery in the sitting and supine positions were analyzed. Telemetric ICP values, wave morphology, and pulse amplitude were determined for functioning and malfunctioning shunts.
Fifty-seven out of 64 patients had available telemetric recordings. The mean ICP was - 3.8 mmHg (standard deviation (SD) = 5.9) in the sitting and 16.4 mmHg (SD = 6.3) in the supine position. The ICP curve demonstrated pulsatility in 49 (86%) patients. A pulsatile curve with mean ICP in the above ranges indicated a functioning shunt, whereas the lack of pulsatility was challenging to interpret. There was a significant positive correlation between ICP versus amplitude, ICP versus body mass index (BMI), and amplitude versus BMI.
This clinical study defined ICP values and curves in IIH patients with a shunt. The results will assist the interpretation of telemetric ICP recordings in clinical decision making. More research is required to model longitudinal recordings and explore the link between telemetric measurements with clinical outcomes.
M.scio 遥测传感器(Aesculap-Miethke,德国)是一种集成在脑室-腹腔(VP)分流器中的设备,用于非侵入性测量颅内压(ICP)。本研究的目的是分析分流的特发性颅内高压(IIH)患者的遥测记录,以确定参考值并协助遥测数据的解释。
这是一项连续队列研究,纳入了 2019 年 7 月至 2022 年 6 月期间因暴发性 IIH 而接受初次 VP 分流术的患者。分析了术后坐位和仰卧位的首次遥测测量。确定了功能和故障分流器的遥测 ICP 值、波形态和脉冲幅度。
64 例患者中有 57 例有可用的遥测记录。坐位时平均 ICP 为-3.8mmHg(标准差(SD)=5.9mmHg),仰卧位时为 16.4mmHg(SD=6.3mmHg)。49 例(86%)患者的 ICP 曲线显示有搏动性。具有上述范围平均 ICP 的搏动性曲线表明分流器功能正常,而缺乏搏动性则难以解释。ICP 与幅度、ICP 与体重指数(BMI)以及幅度与 BMI 之间存在显著正相关。
本临床研究定义了具有分流器的 IIH 患者的 ICP 值和曲线。研究结果将有助于临床决策中对遥测 ICP 记录的解释。需要进一步研究以建立纵向记录模型,并探讨遥测测量与临床结果之间的联系。