Long-term experience in quality assurance of on-rail computed tomography systems for image-guided radiotherapy using in-house multifunctional phantoms.

机构信息

Department of Radiology, University of Yamanashi, 1110 Shimokato, Chuo-City, Yamanashi, 409-3898, Japan.

出版信息

Radiol Phys Technol. 2023 Jun;16(2):292-298. doi: 10.1007/s12194-023-00718-1. Epub 2023 Apr 20.

Abstract

To report the long-term quality assurance (QA) experience of an on-rail computed tomography (CT) system for image-guided radiotherapy using an in-house phantom. An on-rail CT system combining the Elekta Synergy and Canon Aquilion LB was used. The treatment couch was shared by the linear accelerators and CT, and the couch was rotated by 180° when using the on-rail-CT system to ensure that the CT direction was toward the head. All QA analyses were performed by radiation technologists on CBCT or on-rail CT images of the in-house phantom. The CBCT center accuracy from the linac laser, couch rotational accuracy (CBCT center vs. on-rail CT center), horizontal accuracy by CT gantry shift, and remote couch shift accuracy were evaluated. This study reported the QA status of the system during the period 2014-2021. The absolute mean accuracy of couch rotation was 0.4 ± 0.28 mm, 0.44 ± 0.36 mm, and 0.37 ± 0.27 mm in the SI, RL, and AP directions, respectively. Horizontal and remote movement accuracies of the treatment couch were also within 0.5 mm of the absolute mean value. A decrease in the accuracy of couch rotation was also observed due to aging deterioration of related parts caused by the frequent use of couch rotation. The three-dimensional accuracy of on-rail CT systems derived mainly from treatment couches can be maintained within 0.5 mm with appropriate accuracy assurance for at least > 8 years.

摘要

为了报告使用内部体模的轨上 CT 系统进行图像引导放射治疗的长期质量保证 (QA) 经验。使用 Elekta Synergy 和 Canon Aquilion LB 相结合的轨上 CT 系统。治疗床由直线加速器和 CT 共享,当使用轨上 CT 系统时,治疗床旋转 180°,以确保 CT 方向朝向头部。所有 QA 分析均由放射技师在 CBCT 或内部体模的轨上 CT 图像上进行。评估了来自直线加速器激光的 CBCT 中心准确性、治疗床旋转准确性(CBCT 中心与轨上 CT 中心)、通过 CT 龙门架移位的水平准确性和远程治疗床移位准确性。本研究报告了 2014 年至 2021 年期间系统的 QA 状况。治疗床旋转的绝对平均准确性在 SI、RL 和 AP 方向上分别为 0.4±0.28mm、0.44±0.36mm 和 0.37±0.27mm。治疗床的水平和远程移动准确性也在绝对平均值的 0.5mm 以内。由于治疗床旋转的频繁使用导致相关部件的老化恶化,也观察到治疗床旋转准确性的降低。轨上 CT 系统的三维准确性主要来源于治疗床,在适当的精度保证下,至少可保持 8 年以上的 0.5mm 以内。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验