Akagi Takashi, Maeda Takuya, Suga Masaki, Yamashita Tomohiro
Department of Radiation Physics, Hyogo Ion Beam Medical Center, Japan.
Research Division, Hyogo Ion Beam Medical Center, Japan.
Phys Med. 2023 May;109:102587. doi: 10.1016/j.ejmp.2023.102587. Epub 2023 Apr 21.
To evaluate the applicability of microdosimetric kinetic model (MKM) to helium-ion therapy by forming a spread-out Bragg peak (SOBP) of a helium-ion beam using the MKM developed for carbon-ion radiotherapy and confirming the predictions in biological experiments.
Using a ridge filter, a 90-mm wide SOBP for a 210 MeV/u helium-ion beam was created in a broad beam delivery system. The ridge filter was designed such that a uniform biological response was achieved with a cell survival rate of 7% over the SOBP region. Biological experiments were then performed using the SOBP beam in a human salivary gland (HSG) cell line to measure the cell survival rates.
The biological responses were uniform in the SOBP region, as expected by the MKM; however, the mean of the measured cell survival rates was (11.2 ± 0.6) % in the SOBP region, which was 60% higher than the designed rate. When investigating the biological parameters of the HSG cell line used in the experiments, we found that they were altered slightly from the MKM parameters used for carbon-ion radiotherapy. The new β parameter reproduced the measured survival rates within 6.5% in the SOBP region.
We produced biologically uniform SOBP using MKM for carbon-ion radiotherapy. The measured survival rates in the SOBP region were higher than expected, and the survival rates were reproduced by modifying the MKM parameter. This study was limited to one SOBP, and further investigations are required to prove that MKM is generally applicable to helium-ion radiotherapy.
通过使用为碳离子放射治疗开发的微剂量动力学模型(MKM)形成氦离子束的扩展布拉格峰(SOBP),并在生物学实验中确认预测结果,来评估微剂量动力学模型对氦离子治疗的适用性。
在宽束传输系统中,使用脊形滤波器为210 MeV/u的氦离子束创建了一个90毫米宽的SOBP。设计脊形滤波器,以便在SOBP区域内实现均匀的生物学响应,细胞存活率为7%。然后使用SOBP束在人唾液腺(HSG)细胞系中进行生物学实验,以测量细胞存活率。
如MKM所预期的,SOBP区域内的生物学响应是均匀的;然而,SOBP区域内测量的细胞存活率平均值为(11.2±0.6)%,比设计值高60%。在研究实验中使用的HSG细胞系的生物学参数时,我们发现它们与用于碳离子放射治疗的MKM参数略有不同。新的β参数在SOBP区域内将测量的存活率再现误差控制在6.5%以内。
我们使用用于碳离子放射治疗的MKM产生了生物学上均匀的SOBP。SOBP区域内测量的存活率高于预期,通过修改MKM参数再现了存活率。本研究仅限于一个SOBP,需要进一步研究以证明MKM普遍适用于氦离子放射治疗。