Department of Radiation Oncology, Emory University School of Medicine, Atlanta, Georgia, USA.
Mechanical Engineering, Nuclear Radiological Engineering & Medical Physics, Georgia Institute of Technology, Atlanta, Georgia, USA.
J Appl Clin Med Phys. 2023 Sep;24(9):e13997. doi: 10.1002/acm2.13997. Epub 2023 Apr 26.
Improving efficiency of intensity modulated proton therapy (IMPT) treatment can be achieved by shortening the beam delivery time. The purpose of this study is to reduce the delivery time of IMPT, while maintaining the plan quality, by finding the optimal initial proton spot placement parameters.
Seven patients previously treated in the thorax and abdomen with gated IMPT and voluntary breath-hold were included. In the clinical plans, the energy layer spacing (ELS) and spot spacing (SS) were set to 0.6-0.8 (as a scale factor of the default values). For each clinical plan, we created four plans with ELS increased to 1.0, 1.2, 1.4, and SS to 1.0 while keeping all other parameters unchanged. All 35 plans (130 fields) were delivered on a clinical proton machine and the beam delivery time was recorded for each field.
Increasing ELS and SS did not cause target coverage reduction. Increasing ELS had no effect on critical organ-at-risk (OAR) doses or the integral dose, while increasing SS resulted in slightly higher integral and selected OAR doses. Beam-on times were 48.4 ± 9.2 (range: 34.1-66.7) seconds for the clinical plans. Time reductions were 9.2 ± 3.3 s (18.7 ± 5.8%), 11.6 ± 3.5 s (23.1 ± 5.9%), and 14.7 ± 3.9 s (28.9 ± 6.1%) when ELS was changed to 1.0, 1.2, and 1.4, respectively, corresponding to 0.76-0.80 s/layer. SS change had a minimal effect (1.1 ± 1.6 s, or 1.9 ± 2.9%) on the beam-on time.
Increasing the energy layers spacing can reduce the beam delivery time effectively without compromising IMPT plan quality; increasing the SS had no meaningful impact on beam delivery time and resulted in plan-quality degradation in some cases.
通过缩短射束传输时间,可以提高强度调制质子治疗(IMPT)的效率。本研究旨在通过寻找最佳的初始质子点放置参数,来减少 IMPT 的传输时间,同时保持计划质量。
纳入了 7 例先前在胸部和腹部接受门控 IMPT 和自愿屏气治疗的患者。在临床计划中,能量层间距(ELS)和点间距(SS)设置为 0.6-0.8(作为默认值的比例因子)。对于每个临床计划,我们创建了四个计划,将 ELS 增加到 1.0、1.2、1.4,同时将 SS 增加到 1.0,而保持所有其他参数不变。所有 35 个计划(130 个野)均在临床质子机上传输,并记录每个野的射束传输时间。
增加 ELS 和 SS 不会导致靶区覆盖减少。增加 ELS 对关键危及器官(OAR)剂量或积分剂量没有影响,而增加 SS 会导致略高的积分剂量和选定的 OAR 剂量。临床计划的射束开启时间为 48.4±9.2(范围:34.1-66.7)秒。当 ELS 分别改为 1.0、1.2 和 1.4 时,时间减少分别为 9.2±3.3 秒(18.7±5.8%)、11.6±3.5 秒(23.1±5.9%)和 14.7±3.9 秒(28.9±6.1%),相当于 0.76-0.80 秒/层。SS 变化对射束开启时间的影响最小(1.1±1.6 秒,或 1.9±2.9%)。
增加能量层间距可以有效地减少射束传输时间,而不会影响 IMPT 计划质量;增加 SS 对射束传输时间没有明显影响,但在某些情况下会导致计划质量下降。