Molecules. 2023 Apr 13;28(8):3418. doi: 10.3390/molecules28083418.
Herein, we reported an HPLC method for the simultaneous determination of tibezonium iodide (TBN) and lignocaine hydrochloride (LGN). The method was developed according to the International Conference for Harmonization guidelines (ICH) Q2R1 using Agilent 1260 with a mobile phase consisting of acetonitrile and phosphate buffer (pH 4.5) in a volumetric ratio of 70:30 and flowing through a C Agilent column at 1 mL/min. The results revealed that TBN and LGN peaks were isolated at 4.20 and 2.33 min, respectively, with a resolution of 2.59. The accuracy of TBN and LGN was calculated to be 100.01 ± 1.72% and 99.05 ± 0.65% at 100% concentration, respectively. Similarly, the respective precision was 100.03 ± 1.61% and 99.05 ± 0.48%. The repeatability for TBN and LGN was found to be 99.05 ± 0.48% and 99.19 ± 1.72%, respectively, indicating that the method was precise. The respective regression co-efficient (r) for TBN and LGN was found to be 0.9995 and 0.9992. Moreover, the LOD and LOQ values for TBN were 0.012 and 0.037 µg/mL, respectively, while for LGN, they were 0.115 and 0.384 µg/mL, respectively. The calculated greenness of the method for ecological safety was found to be 0.83, depicting a green contour on the AGREE scale. No interfering peaks were found when the analyte was estimated in dosage form and in volunteers' saliva, depicting the specificity of the method. Conclusively, a robust, fast, accurate, precise and specific method was successfully validated to estimate TBN and LGN.
本文报道了一种同时测定碘替硝唑(TBN)和盐酸利多卡因(LGN)的 HPLC 方法。该方法是根据国际协调会议(ICH)Q2R1 指导原则建立的,采用 Agilent 1260 高效液相色谱仪,流动相为乙腈-磷酸盐缓冲液(pH4.5)(70:30),流速为 1mL/min,色谱柱为 Agilent C 柱。结果表明,TBN 和 LGN 的色谱峰分别在 4.20 和 2.33min 处分离,分离度为 2.59。TBN 和 LGN 的准确度分别计算为 100.01±1.72%和 99.05±0.65%,浓度为 100%时。同样,精密度分别为 100.03±1.61%和 99.05±0.48%。TBN 和 LGN 的重复性分别为 99.05±0.48%和 99.19±1.72%,表明该方法精密度良好。TBN 和 LGN 的回归系数(r)分别为 0.9995 和 0.9992。此外,TBN 的 LOQ 和 LOD 值分别为 0.012 和 0.037μg/mL,而 LGN 的 LOQ 和 LOD 值分别为 0.115 和 0.384μg/mL。该方法的生态安全性绿色度计算值为 0.83,在 AGREE 量表上呈现绿色轮廓。在剂型和志愿者唾液中测定时,未发现干扰峰,表明该方法具有特异性。结论:成功验证了一种用于同时测定 TBN 和 LGN 的稳健、快速、准确、精密和专属性方法。