Linker D T, Moritz W E, Pearlman A S
J Am Coll Cardiol. 1986 Jul;8(1):101-6. doi: 10.1016/s0735-1097(86)80098-5.
Right ventricular volume is difficult to measure accurately from one or two views because the complexity of right ventricular shape invalidates simplifying geometric assumptions. This article describes a new three-dimensional echocardiographic reconstruction method of right ventricular volume calculation, and reports the results of testing this method in vitro using normal animal hearts and pathologic specimens from infants and children who died with aortic stenosis or hypoplastic left heart. The correlation with reference volumes was excellent for both groups (r = 0.98, n = 25 for the animal data; r = 0.97, n = 15 for the human data). Given the calculated echocardiographic volume (Vc), the reference volume (Vr) was best estimated by the equation Vr = 1.16 Vc for the animal data and Vr = 0.92 Vc for the human data. Three-dimensional echocardiographic measurement of right ventricular volume is an accurate method that deserves further study and application in a clinical setting.
由于右心室形状复杂,使得简化的几何假设无效,因此仅从一两个视图很难准确测量右心室容积。本文描述了一种计算右心室容积的新型三维超声心动图重建方法,并报告了使用正常动物心脏以及患有主动脉狭窄或左心发育不全而死亡的婴幼儿病理标本对该方法进行体外测试的结果。两组的测量结果与参考容积的相关性都非常好(动物数据:r = 0.98,n = 25;人类数据:r = 0.97,n = 15)。根据计算得到的超声心动图容积(Vc),动物数据的参考容积(Vr)最佳估计方程为Vr = 1.16 Vc,人类数据为Vr = 0.92 Vc。三维超声心动图测量右心室容积是一种准确的方法,值得在临床环境中进一步研究和应用。