Vrije Universiteit Brussel (VUB), UZ Brussel, KidZ Health Castle, Brussels, Belgium.
Vrije Universiteit Brussel (VUB), UZ Brussel, KidZ Health Castle, Brussels, Belgium.
J Pediatr. 2023 Aug;259:113449. doi: 10.1016/j.jpeds.2023.113449. Epub 2023 May 6.
To assess the reliability of the KidZ Health Castle formula (KHC-F) to determine the correct probe position of a multichannel intraluminal impedance pH.
A retrospective cohort study was performed on 222 children between 1 month and 18 years of age undergoing multichannel intraluminal impedance pH. The primary outcome was the comparison of the pH sensor location determined by the KHC-F with the radiological target position. The margin of error was defined as 1 cm from the target position. Performance of the KHC-F and existing formulas was determined via the percentage with a correct position, mean error, 95% limits of agreement (Bland-Altman plots), and Spearman correlation. A post hoc analysis was performed with an updated KHC-F v2, subtracting -0.5 cm from the KHC-F.
Positioning with KHC-F was correct in two-thirds of the participants, with a very strong correlation (ρ = 0.91) with the target position. Bland-Altman plots showed good agreement between KHC-F and target position (mean error of -0.44 cm, lower limit -3.2 cm, upper limit 2.3 cm). A post hoc analysis with the KHC-F v2 showed a correct positioning in 74% of patients. Comparison with other formulas showed a stronger performance of KHC-F and KHC-F v2 on correct positioning, mean error, and 95% limits of agreement.
The KHC-F leads to reliable results. KHC-F v2 outperforms all other existing formulas in children, thereby reducing the need for repositioning and the amount of x-ray exposure. The age distribution of the sample may be a limitation, as well as the retrospective nature of the study.
评估 KidZ Health Castle 公式(KHC-F)确定多通道腔内阻抗 pH 正确探头位置的可靠性。
对 222 名 1 个月至 18 岁行多通道腔内阻抗 pH 检查的儿童进行回顾性队列研究。主要结局是比较 KHC-F 确定的 pH 传感器位置与放射学靶位置。误差范围定义为距离靶位置 1 cm。通过正确位置的百分比、平均误差、95%一致性界限(Bland-Altman 图)和 Spearman 相关系数来确定 KHC-F 和现有公式的性能。使用更新的 KHC-F v2 进行了事后分析,从 KHC-F 中减去-0.5 cm。
KHC-F 定位在三分之二的参与者中是正确的,与靶位置具有很强的相关性(ρ=0.91)。Bland-Altman 图显示 KHC-F 与靶位置之间具有良好的一致性(平均误差为-0.44 cm,下限为-3.2 cm,上限为 2.3 cm)。KHC-F v2 的事后分析显示,74%的患者定位正确。与其他公式的比较表明,KHC-F 和 KHC-F v2 在正确定位、平均误差和 95%一致性界限方面具有更强的性能。
KHC-F 可得出可靠的结果。KHC-F v2 在儿童中优于所有其他现有公式,从而减少了重新定位和 X 射线照射的次数。样本的年龄分布可能是一个限制,以及研究的回顾性性质。