Department of Obstetrics and Gynecology, The First Affiliated Hospital, Fujian Medical University, Fuzhou 350005, China.
Department of Obstetrics and Gynecology, National Regional Medical Center, Binhai Campus of the First Affiliated Hospital, Fujian Medical University, Fuzhou 350212, China.
Anal Chem. 2023 May 23;95(20):8121-8127. doi: 10.1021/acs.analchem.3c01246. Epub 2023 May 11.
In this work, a simple and sensitive electrochemiluminescence (ECL) biosensor has been devised based on target-induced steric hindrance of an antibody-modified electrode surface. Estrogen-related receptor alpha (ERRα) is closely related to estrogen-dependent tumors, which had been chosen as a model target. The ERRα antigen can bind to the antibody modified on the electrode surface with high specificity and results in the increase of steric hindrance, which prevented the ECL indicators (tris(2,2'-bipyridine) dichlororuthenium(II) hexahydrate) from approaching the electrode surface, and the ECL intensity of the system decreased. The ECL response of the system has a good linear relationship with ERRα concentration in the range of 1.0-60 ng/L, and the limit of detection is 0.5 ng/L. Different from the traditional sandwiched immune ECL detection system, which need the modification of ECL indicators on the secondary antibody, only one antibody had been used in this system. The system is easy to operate and has good sensitivity. The designed biosensor has been applied to detect ERRα in the serum and different cell line samples with satisfied results.
在这项工作中,设计了一种基于抗体修饰电极表面的目标诱导空间位阻的简单灵敏的电化学发光(ECL)生物传感器。雌激素相关受体α(ERRα)与雌激素依赖性肿瘤密切相关,已被选为模型靶标。ERRα 抗原可以与电极表面修饰的抗体特异性结合,导致空间位阻增加,阻止 ECL 指示剂(六水合二氯二(2,2'-联吡啶)钌(II))接近电极表面,从而降低了该体系的 ECL 强度。该体系的 ECL 响应与 ERRα 浓度在 1.0-60ng/L 范围内呈良好的线性关系,检出限为 0.5ng/L。与传统的夹心免疫 ECL 检测系统不同,该系统需要在二级抗体上修饰 ECL 指示剂,而该系统只使用了一种抗体。该系统操作简单,灵敏度高。该设计的生物传感器已应用于检测血清和不同细胞系样本中的 ERRα,结果令人满意。