This paper presents a model-based approach for the design of the primary drying stage of a freeze-drying process using a small-scale freeze-dryer (MicroFD® by Millrock Technology Inc.). Gravimetric tests, coupled with a model of the heat transfer to the product in the vials that account also for the heat exchange between the edge vials and the central vials, are used to infer the heat transfer coefficient from the shelf to the product in the vial (K), that is expected to be (almost) the same in different freeze-dryers. Differently from other approaches previously proposed, the operating conditions in MicroFD® are not chosen to mimic the dynamics of another freeze-dryer: this allows saving time and resources as no experiments are needed in the large-scale unit, and no additional tests in the small-scale unit, apart from the three gravimetric tests usually needed to assess the effect of chamber pressure on K. With respect to the other model parameter, R, the resistance of the dried cake to mass transfer, it is not influenced by the equipment and, thus values obtained in a freeze-dryer may be used to simulate the drying in a different unit, provided the same filling conditions are used, as well as the same operating conditions in the freezing stage, and cake collapse (or shrinkage) is avoided. The method was validated considering ice sublimation in two types of vials (2R and 6R) and at different operating conditions (6.7, 13.3 and 26.7 Pa), with the freeze-drying of a 5% w/w sucrose solution as a test case. An accurate estimate for both K and R was obtained with respect to the values obtained in a pilot-scale equipment, determined through independent tests for validation purposes. Simulation of the product temperature and drying time in a different unit was then possible, and results were validated experimentally.
本文提出了一种基于模型的方法,用于使用小型冷冻干燥器(Millrock Technology Inc. 的 MicroFD®)设计冷冻干燥过程的初级干燥阶段。通过重量测试,结合小瓶中产品传热的模型,该模型还考虑了边缘小瓶和中心小瓶之间的热交换,从搁板到小瓶中产品的传热系数(K)可以推断出来,预计在不同的冷冻干燥器中(几乎)相同。与之前提出的其他方法不同,MicroFD®中的操作条件不是为了模拟另一个冷冻干燥器的动态而选择的:这可以节省时间和资源,因为在大型设备中不需要进行实验,在小型设备中也不需要进行额外的测试,除了通常需要进行的三个重量测试以评估腔室压力对 K 的影响。关于另一个模型参数 R,即干燥蛋糕对质量传递的阻力,它不受设备的影响,因此可以在冷冻干燥器中获得的值用于模拟不同单元中的干燥,只要使用相同的填充条件,以及在冷冻阶段使用相同的操作条件,并避免蛋糕塌陷(或收缩)。该方法通过在两种类型的小瓶(2R 和 6R)和不同的操作条件(6.7、13.3 和 26.7 Pa)下考虑冰升华进行了验证,以 5%w/w 蔗糖溶液的冷冻干燥作为测试案例。相对于通过独立测试确定的用于验证目的的中试设备中获得的值,对 K 和 R 进行了准确估计。然后可以模拟不同单元中的产品温度和干燥时间,并通过实验验证结果。