Anastomotic leaks (AL) and staple line leaks are a serious post-operative complication that can develop following bariatric surgery. The delay in the onset of symptoms following a leak usually results in reactive diagnostics and treatment, leading to increased patient morbidity and mortality, and a clinical and economic burden on both the patient and the hospital. Despite support in literature for pH as a biomarker for early detection of AL, the current methods of pH detection require significant clinician involvement and resources. Presented here is a polyaniline (PANI)-based pH sensor that can be connected inline to surgical drains to continuously monitor peritoneal secretion in real time for homeostatic changes in pH. During this study, the baseline peritoneal fluid pH was measured in two pigs using the PANI sensor and verified using a benchtop pH probe. The PANI sensor was then utilized to continuously monitor the changes in the pH of peritoneal effluent, as a gastric leak was simulated. The inline sensors were able to detect the resulting local changes in drainage pH within 10 min of leak induction. The successful implementation of this sensor in clinical practice can both enable high efficiency continuous monitoring of patient status and drastically decrease the time required to detect AL, thus potentially decreasing the clinical and economic burden incurred by gastric leaks.
吻合口漏(AL)和吻合钉线漏是减肥手术后可能出现的严重术后并发症。漏出后症状出现的延迟通常会导致反应性诊断和治疗,从而增加患者的发病率和死亡率,并给患者和医院带来临床和经济负担。尽管文献支持将pH值作为早期检测AL的生物标志物,但目前的pH检测方法需要临床医生大量参与并耗费资源。本文介绍了一种基于聚苯胺(PANI)的pH传感器,该传感器可以在线连接到手术引流管,以实时连续监测腹膜分泌物的pH值稳态变化。在这项研究中,使用PANI传感器在两只猪身上测量了基线腹膜液pH值,并使用台式pH探头进行了验证。然后,在模拟胃漏的情况下,利用PANI传感器连续监测腹膜流出液pH值的变化。在线传感器能够在漏出诱导后10分钟内检测到引流液pH值的局部变化。该传感器在临床实践中的成功应用既能实现对患者状态的高效连续监测,又能大幅减少检测AL所需的时间,从而有可能减轻胃漏带来的临床和经济负担。