Newborn Services, John Radcliffe Hospital, Oxford University Hospital, Oxford, UK; Division of Neonatology, Department of Woman and Child, ASST Settelaghi, University of Insubria, Varese, Italy.
NICU "V. Buzzi" Children's Hospital, ASST FBF-Sacco, Milan, Italy.
Early Hum Dev. 2023 Aug;183:105794. doi: 10.1016/j.earlhumdev.2023.105794. Epub 2023 Jun 1.
Infants <28 weeks' gestation in need of inflations at birth were recorded with Respiratory Function Monitor. Two devices were used for resuscitation. Peak Inspiratory Pressure spikes were visible in all inflations with GE Panda and in none with Neo-Puff. There was no significant difference in mean Vte/kg between GE Panda and Neo-Puff.