Suppr超能文献

Kinetic fluorometric determination of aluminum in serum.

作者信息

Ioannou P C, Piperaki E A

出版信息

Clin Chem. 1986 Aug;32(8):1481-3.

PMID:3731441
Abstract

We describe a simple fluorometric method for determining aluminum in serum samples by monitoring the rate of reaction of 2-hydroxy-1-naphthaldehyde-p-methoxybenzoylhydrazone with aluminum ions. The emission of the resulting fluorescent metal-chelate formed is measured at 475 nm. Aluminum was measured in the supernate of serum after proteins were removed by precipitation with concentrated nitric acid, and calculations were based on the technique of standard additions. Within-run precision (CV) was 7.8% and 4.8% at mean aluminum concentrations of 7.7 and 60.7 micrograms/L, respectively (n = 10); between-run precision (CV) was 8.9% and 5.7% at mean aluminum concentrations of 23.3 and 46.8 micrograms/L, respectively (n = 10). The standard curve for the method is linear over the range of 0-250 micrograms of aluminum per liter. Samples from 49 patients were analyzed for aluminum by the proposed method (y) and by electrothermal atomic absorption spectroscopy (x). Linear regression analysis of the results yielded the equation y = 0.98x + 2.3 (r = 0.989, Syx = 6.7). The proposed method is comparable in sensitivity to the well-accepted atomic absorption spectrometric method but is simpler and less expensive.