Orthopaedics, Unit for Surgical Reconstruction of Major Joints, University General Hospital of Alexandroupolis, Alexandroupolis, Greece.
Orthopaedics, Unit for Surgical Reconstruction of Major Joints, University General Hospital of Alexandroupolis, Alexandroupolis, Greece.
Med Eng Phys. 2023 Jul;117:103991. doi: 10.1016/j.medengphy.2023.103991. Epub 2023 May 18.
The number of total knee arthroplasties (TKA) is rising in many countries along with the numbers of the revision TKA. Rotating hinge knee (RHK) implants have a fundamental role in the revision TKA and their design has evolved during the last years becoming appealing to many surgeons worldwide. They are mainly used in cases where large bone defects and severe soft tissue imbalance exist. However, despite their recent advancements, they are still associated with high complication rates such as infection, periprosthetic fractures and insufficiency of the extensor apparatus. An uncommon complication of the latest rotating hinge implants is their mechanical component failure. Here we present a rare case of such a devastating complication where a modern RHK prosthesis dislocated without a preceding traumatic event, a review of the corresponding literature and a potential cause for the failure of the mechanism. Moreover, insight on key aspects that need to be addressed is provided, like intrinsic and extrinsic factors which play a crucial role and shouldn't be overlooked for a successful outcome.
随着全膝关节置换术(TKA)数量的增加,许多国家的翻修 TKA 数量也在增加。旋转铰链膝关节(RHK)植入物在翻修 TKA 中起着重要作用,其设计在过去几年中不断发展,受到全球许多外科医生的青睐。它们主要用于存在大骨缺损和严重软组织失衡的情况。然而,尽管最近取得了进展,但它们仍然与高并发症发生率相关,如感染、假体周围骨折和伸肌装置不足。最新的旋转铰链植入物的一种罕见并发症是其机械部件故障。在这里,我们介绍了一个罕见的灾难性并发症的案例,一个现代的 RHK 假体在没有先前创伤事件的情况下脱位,对相应文献的回顾以及该机制失效的潜在原因。此外,还提供了对需要解决的关键方面的深入了解,例如内在和外在因素,它们起着至关重要的作用,不容忽视,以获得成功的结果。