Zhuang Tingliang, Gibbard Grant, Duan Xinhui, Tan Jun, Park Yang, Lin Mu-Han, Sun Zhihui, Oderinde Oluwaseyi M, Lu Weiguo, Reynolds Robert, Godley Andrew, Pompos Arnold, Dan Tu, Garant Aurelie, Iyengar Puneeth, Timmerman Robert, Jiang Steve, Cai Bin
Department of Radiation Oncology, University of Texas- Southwestern Medical Center, Dallas, USA.
Department of Radiology, University of Texas- Southwestern Medical Center, Dallas, USA.
Phys Imaging Radiat Oncol. 2023 Apr 10;26:100438. doi: 10.1016/j.phro.2023.100438. eCollection 2023 Apr.
A recently developed biology-guided radiotherapy platform, equipped with positron emission tomography (PET) and computed tomography (CT), provides both anatomical and functional image guidance for radiotherapy. This study aimed to characterize performance of the kilovoltage CT (kVCT) system on this platform using standard quality metrics measured on phantom and patient images, using CT simulator images as reference.
Image quality metrics, including spatial resolution/modular transfer function (MTF), slice sensitivity profile (SSP), noise performance and image uniformity, contrast-noise ratio (CNR) and low-contrast resolution, geometric accuracy, and CT number (HU) accuracy, were evaluated on phantom images. Patient images were evaluated mainly qualitatively.
On phantom images the MTF is about 0.68 lp/mm for kVCT in PET/CT Linac. The SSP agreed with nominal slice thickness within 0.7 mm. The diameter of the smallest visible target (1% contrast) is about 5 mm using medium dose mode. The image uniformity is within 2.0 HU. The geometric accuracy tests passed within 0.5 mm. Relative to CT simulator images, the noise is generally higher and the CNR is lower in PET/CT Linac kVCT images. The CT number accuracy is comparable between the two systems with maximum deviation from the phantom manufacturer range within 25 HU. On patient images, higher spatial resolution and image noise are observed on PET/CT Linac kVCT images.
Major image quality metrics of the PET/CT Linac kVCT were within vendor-recommended tolerances. Better spatial resolution but higher noise and better/comparable low contrast visibility were observed as compared to a CT simulator when images were acquired with clinical protocols.
最近开发的一种结合正电子发射断层扫描(PET)和计算机断层扫描(CT)的生物引导放射治疗平台,可为放射治疗提供解剖和功能图像引导。本研究旨在使用在体模和患者图像上测量的标准质量指标,以CT模拟图像作为参考,来表征该平台上千伏CT(kVCT)系统的性能。
在体模图像上评估图像质量指标,包括空间分辨率/调制传递函数(MTF)、层厚灵敏度曲线(SSP)、噪声性能和图像均匀性、对比噪声比(CNR)和低对比度分辨率、几何精度以及CT值(HU)精度。对患者图像主要进行定性评估。
在PET/CT直线加速器中的kVCT,体模图像上的MTF约为0.68 lp/mm。SSP与标称层厚的偏差在0.7 mm以内。使用中等剂量模式时,最小可见目标(1%对比度)的直径约为5 mm。图像均匀性在2.0 HU以内。几何精度测试在0.5 mm以内通过。相对于CT模拟图像,PET/CT直线加速器kVCT图像中的噪声通常更高,CNR更低。两个系统的CT值精度相当,与体模制造商规定范围的最大偏差在25 HU以内。在患者图像上,PET/CT直线加速器kVCT图像具有更高的空间分辨率和图像噪声。
PET/CT直线加速器kVCT的主要图像质量指标在供应商推荐的公差范围内。与CT模拟器相比,当按照临床方案采集图像时,观察到更好的空间分辨率,但噪声更高,低对比度可见性更好/相当。