Rangel Jonhattan Ferreira, de Almeida Santos Willyam Brito, de Carvalho Costa Thércio Henrique, de Bessa Kleiber Lima, Melo José Daniel Diniz
Materials Science and Engineering Post-Graduation, Department of Materials Science and Engineering, Federal University of Rio Grande do Norte-UFRN, Natal 59078-970, Brazil.
Mechanical Post-Graduation, Department of Mechanical Engineering, Federal University of Rio Grande do Norte-UFRN, Natal 59078-970, Brazil.
J Funct Biomater. 2023 Jun 2;14(6):310. doi: 10.3390/jfb14060310.
An arteriovenous fistula (AVF) is the access most recommended by several authors. However, its manufacture and use can cause several problems in the short, medium and long term. The study of fluid dynamics related to the structure of the AVF can provide information necessary for the reduction of these problems and a better quality of life for patients. The present study analyzed pressure variation in a rigid and flexible (thickness variation) model of AVFs manufactured based on patient data. A computed tomography was performed from which the geometry of the AVF was removed. This was treated and adapted to the pulsatile flow bench. Bench tests with simulation of systolic-diastolic pulse showed higher pressure peaks in the rigid AVF followed by the flexible model with 1 mm thickness. The inflection of the pressure values of the flexible AVF in relation to the rigid one was observed, being more expressive in the flexible AVF of 1 mm. The 1 mm flexible AVF presented an average pressure close to the physiological one and a smaller pressure drop, showing that this AVF model presents the best condition among the three to serve as a basis for the development of an AVF substitute.
动静脉内瘘(AVF)是多位作者最为推荐的血管通路。然而,其制作和使用在短期、中期和长期都可能引发若干问题。对与AVF结构相关的流体动力学进行研究,可为减少这些问题及提高患者生活质量提供必要信息。本研究分析了基于患者数据制作的刚性和柔性(厚度可变)AVF模型中的压力变化。进行了计算机断层扫描,从中提取了AVF的几何形状。对其进行处理并适配到脉动流实验台上。模拟收缩 - 舒张脉冲的台架试验表明,刚性AVF中的压力峰值更高,其次是厚度为1毫米的柔性模型。观察到柔性AVF相对于刚性AVF的压力值拐点,在1毫米的柔性AVF中更为明显。1毫米的柔性AVF呈现出接近生理压力的平均压力和较小的压力降,表明该AVF模型在三者中呈现出作为AVF替代物开发基础的最佳条件。