Leonard R L, Bull A B, Xue F, Haycook C P, Gray S K, Bond C W, Bond P E, McDearman J C, Woods D P, Mayfield J, Brown L R, Giorgio T D, Johnson J A
Department of Mechanical, Aerospace, and Biomedical Engineering, University of Tennessee Space Institute, Tullahoma, TN 37388.
Center for Laser Applications, University of Tennessee Space Institute, Tullahoma, TN 37388.
Appl Surf Sci. 2023 Oct 15;634. doi: 10.1016/j.apsusc.2023.157606. Epub 2023 May 27.
Laparoscopes can suffer from fogging and contamination difficulties, resulting in a reduced field of view during surgery. A series of diamond-like carbon films, doped with SiO, were produced by pulsed laser deposition for evaluation as biocompatible, antifogging coatings. DLC films doped with SiO demonstrated hydrophilic properties with water contact angles under 40°. Samples subjected to plasma cleaning had improved contact angle results, with values under 5°. Doping the DLC films with SiO led to an average 40% decrease in modulus and 60% decrease in hardness. Hardness of the doped films, 12.0 - 13.2 GPa, was greater than that of the uncoated fused silica substrate, 9.2 GPa. The biocompatibility was assessed through CellTiter-Glo assays, with the films demonstrating statistically similar levels of cell viability when compared to the control media. The absence of ATP released by blood platelets in contact with the DLC coatings suggests hemocompatibility. The SiO doped films displayed improved transparency levels in comparison to undoped films, achieving up to an average of 80% transmission over the visible spectrum and an attenuation coefficient of 1.1 × 10 cm at the 450 nm wavelength. The SiO doped DLC films show promise as a method of fog prevention for laparoscopes.
腹腔镜可能会出现起雾和污染问题,导致手术过程中的视野受限。通过脉冲激光沉积制备了一系列掺杂SiO的类金刚石碳膜,用于评估其作为生物相容性防雾涂层的性能。掺杂SiO的类金刚石碳膜表现出亲水性,水接触角低于40°。经过等离子体清洗的样品接触角结果有所改善,角度值低于5°。在类金刚石碳膜中掺杂SiO导致模量平均降低40%,硬度降低60%。掺杂膜的硬度为12.0 - 13.2 GPa,高于未涂层的熔融石英基底的硬度9.2 GPa。通过CellTiter-Glo检测评估生物相容性,与对照培养基相比,这些膜显示出统计学上相似的细胞活力水平。与类金刚石碳涂层接触的血小板未释放ATP,表明具有血液相容性。与未掺杂的膜相比,掺杂SiO的膜透明度有所提高,在可见光谱范围内平均透过率高达80%,在450 nm波长处的衰减系数为1.1×10 cm。掺杂SiO的类金刚石碳膜有望成为一种腹腔镜防雾方法。