Machin G, Johnson B Carol, Gibson C, Rusby R L
The National Physical Laboratory, Queens Road, Teddington, Middlesex, TW11 OLW, United Kingdom.
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001, USA.
J Res Natl Inst Stand Technol. 1994 Nov-Dec;99(6):731-736. doi: 10.6028/jres.099.065.
An intercomparison of radiance temperature scales has been performed by the National Physical Laboratory (NPL) and the National Institute of Standards and Technology (NIST) using a standard transfer pyrometer operating at a wavelength of approximately 1000 nm. It was found that the radiance temperature scales established by the two laboratories were in agreement to 0.1% or better of the temperature over the range 1000 °C to 2500 °C.
英国国家物理实验室(NPL)和美国国家标准与技术研究院(NIST)使用一台工作波长约为1000纳米的标准传递高温计,对辐射温度量表进行了比对。结果发现,两个实验室建立的辐射温度量表在1000℃至2500℃的温度范围内,其一致性达到温度的0.1%或更佳。