Suppr超能文献

丹纳美(Dinamap)血压计无法检测出极低出生体重儿的低血压情况。

Dinamap fails to detect hypotension in very low birthweight infants.

作者信息

Diprose G K, Evans D H, Archer L N, Levene M I

出版信息

Arch Dis Child. 1986 Aug;61(8):771-3. doi: 10.1136/adc.61.8.771.

Abstract

The accuracy of blood pressure measurements obtained from very low birthweight (less than 1500 g) neonates using the Dinamap oscillometric monitor was investigated. Comparisons using umbilical artery measurements showed that the monitor is less reliable in the lower pressure range and specifically that it tends to overestimate pressure in hypotensive infants.

摘要

研究了使用Dinamap振荡式监护仪测量极低出生体重(小于1500克)新生儿血压的准确性。与脐动脉测量结果的比较表明,该监护仪在较低血压范围内可靠性较低,特别是在低血压婴儿中往往会高估血压。

相似文献

2
Arterial blood pressure in the very low birthweight neonate. Evaluation of an automatic oscillometric technique.
Acta Paediatr Scand. 1987 Mar;76(2):338-41. doi: 10.1111/j.1651-2227.1987.tb10471.x.

引用本文的文献

3
Advances in Non-Invasive Blood Pressure Monitoring.无创血压监测的进展。
Sensors (Basel). 2021 Jun 22;21(13):4273. doi: 10.3390/s21134273.
7
Accuracy of non-invasive blood pressure monitoring in very preterm infants.非常早产儿无创血压监测的准确性。
Intensive Care Med. 2012 Apr;38(4):670-6. doi: 10.1007/s00134-012-2499-y. Epub 2012 Mar 6.

本文引用的文献

6
Clinical evaluation of the Dinamap non-invasive blood pressure monitor in pre-term neonates.
Clin Phys Physiol Meas. 1981 Nov;2(4):271-6. doi: 10.1088/0143-0815/2/4/003.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验