Levinson M M, Smith R G, Cork R C, Gallo J, Emery R W, Icenogle T B, Ott R A, Burns G L, Copeland J G
Artif Organs. 1986 Jun;10(3):236-44. doi: 10.1111/j.1525-1594.1986.tb02552.x.
The Jarvik-7 total artificial heart (TAH) has now been implanted in seven patients, with four (57%) suffering cerebral embolic events. Functional recovery in two patients was impaired by subsequent intracranial hemorrhage from anticoagulant therapy. The findings of a Jarvik-7 TAH explanted following a cerebral embolus are reported here. Thrombotic deposition was seen in the inner crevice formed between the two components of the aortic quick connector. In addition, deposition was seen around all four valves in the crevice formed between the valve ring and the plastic valve housing. On the atrioventricular valves, the deposition in this crevice was most severe on the outflow surface adjacent to the minor flow orifice. On the pulmonic valve, deposition was seen around the entire circumference of both the inflow and the outflow crevices. No thrombi were seen on the diaphragm, the diaphragm-housing junction, or the housing of the left ventricle. Hypercoagulability of the patient's blood gradually developed during implant, reaching a maximum on the evening before the clinical embolus. Increasing resistance to heparin was also demonstrated by in vitro testing.
贾维克-7型全人工心脏(TAH)现已植入7名患者体内,其中4名(57%)发生了脑栓塞事件。两名患者的功能恢复因抗凝治疗引发的颅内出血而受损。本文报告了一枚在发生脑栓塞后被取出的贾维克-7型TAH的检查结果。在主动脉快速连接器两个部件之间形成的内部缝隙中可见血栓沉积。此外,在瓣膜环与塑料瓣膜外壳之间形成的缝隙中,所有四个瓣膜周围均可见沉积。在房室瓣上,该缝隙中的沉积在靠近小流量孔口的流出表面最为严重。在肺动脉瓣上,流入和流出缝隙的整个圆周周围均可见沉积。在隔膜、隔膜-外壳连接处或左心室外壳上未发现血栓。患者血液的高凝状态在植入过程中逐渐发展,在临床栓塞前一晚达到最高值。体外测试也显示对肝素的抵抗性增加。