Department of Emergency Medicine, Basaksehir Cam and Sakura City Hospital, Istanbul, Turkey.
Department of Biostatistics, İzmir Katip Celebi University Faculty of Medicine, İzmir, Turkey.
Ir J Med Sci. 2024 Feb;193(1):477-483. doi: 10.1007/s11845-023-03456-w. Epub 2023 Jul 13.
Smartwatches have gained tremendous attention in recent years and have become widely accepted by patients, despite not being intended for medical diagnosis.
This study aimed to determine the accuracy of Apple Watch oxygen saturation measurement in patients with acute exacerbation of COPD by comparing it with medical-grade pulse oximetry and ABG.
This single-center, prospective, cross-sectional study involved 167 patients. Patients presenting with cardiac arrest, life-threatening symptoms, severe hypoxia, or obvious jaundice were excluded. Additionally, patients whose SpO measurements with the Apple Watch took more than 2 min or required eight attempts were also excluded. Vital signs were measured simultaneously using the IntelliVue MX500 monitor with the Masimo Rainbow Set pulse oximeter and the Apple Watch. Concurrently, arterial blood gas (ABG) samples were drawn.
A strong correlation between the Apple Watch 6 and medical-grade pulse oximetry (r = 0.89, ICC = 0.940) was noted. The Bland-Altman analysis revealed a mean error of 0.458% between the Apple Watch 6 and ABG (SD: 2.78, level of agreement: - 5.912 to 4.996). The mean error between pulse oximetry and ABG (SD: 5.086, level of agreement; - 10.983 to 8.953) was 1.015%. There was a correlation between respiratory rate and the number of attempts to measure SpO with the Apple Watch 6 (r = 0.75, p < 0.05).
Apple Watch 6 is an accurate and reliable method for measuring SpO levels in emergency patients who presented with acute exacerbation of COPD. However, tachypneic patients may encounter challenges due to the potential need for multiple attempts to measure their oxygen saturation.
近年来,智能手表受到了极大的关注,并得到了患者的广泛认可,尽管它们并非用于医疗诊断。
本研究旨在通过与医疗级脉搏血氧仪和动脉血气(ABG)比较,确定 Apple Watch 在急性加重期 COPD 患者中的血氧饱和度测量准确性。
这是一项单中心、前瞻性、横断面研究,共纳入 167 名患者。排除出现心搏骤停、危及生命的症状、严重缺氧或明显黄疸的患者。此外,排除 Apple Watch 测量 SpO2 超过 2 分钟或需要尝试 8 次的患者。使用 IntelliVue MX500 监护仪和 Masimo Rainbow Set 脉搏血氧仪以及 Apple Watch 同时测量生命体征,并抽取动脉血气样本。
Apple Watch 6 与医疗级脉搏血氧仪之间存在很强的相关性(r=0.89,ICC=0.940)。Bland-Altman 分析显示,Apple Watch 6 与 ABG 之间的平均误差为 0.458%(SD:2.78,一致性水平:-5.912 至 4.996)。脉搏血氧仪与 ABG 之间的平均误差为 1.015%(SD:5.086,一致性水平:-10.983 至 8.953)。呼吸频率与使用 Apple Watch 6 测量 SpO2 的尝试次数之间存在相关性(r=0.75,p<0.05)。
Apple Watch 6 是一种准确可靠的方法,可用于测量急性加重期 COPD 患者的 SpO2 水平。然而,呼吸急促的患者可能会因需要多次尝试测量其血氧饱和度而遇到挑战。