Lee Han-Na, Kim Myoung-Sub, Lee Jeong-Yol, Zihan Xu, Ryu Jae-Jun, Shim Ji-Suk
Department of Dentistry, Korea University Guro Hospital, Seoul, Republic of Korea.
Department of Dentistry, Korea University Anam Hospital, Seoul, Republic of Korea.
J Adv Prosthodont. 2023 Jun;15(3):126-135. doi: 10.4047/jap.2023.15.3.126. Epub 2023 Jun 28.
The aim of this study was to evaluate the reliability of implant stability measuring devices depending on the location of the implant and the position of the patient.
Six implants were installed in different dentate sextants of six artificial bone models. Implant stability was measured in three conditions of the bone model (without mounting on a phantom head, mounted on a phantom head in supine position, and mounted on a phantom head in upright position). A resonance frequency analysis device (Osstell) and two damping capacity analysis devices (Periotest and Anycheck) were used to measure implant stability. The values measured outside the phantom head were treated as controls, and the values inside the phantom head were compared using an independent -test.
Osstell showed different results in two of the six divisions in both the supine and upright positions compared to outside of the mouth ( < .05). Periotest showed different results in all six parts in the supine position and in five parts in the upright position compared to outside of the mouth ( < .05). While Anycheck showed different results in five areas in the supine position compared to outside of the mouth, it showed different results in only one area in the upright position ( < .05).
In the difficult implant position for the operator to access, the implant stability measuring devices show less reliability. The accessibility of implant is greatly affected in the order of Osstell, Anycheck, and Periotest.
本研究的目的是根据种植体的位置和患者的体位评估种植体稳定性测量装置的可靠性。
在六个人工骨模型的不同牙列象限中植入六颗种植体。在骨模型的三种状态下测量种植体稳定性(未安装在仿头模上、仰卧位安装在仿头模上、直立位安装在仿头模上)。使用共振频率分析装置(Osstell)和两种阻尼能力分析装置(Periotest和Anycheck)测量种植体稳定性。将在仿头模外测量的值作为对照,使用独立t检验比较仿头模内的值。
与口腔外相比,Osstell在仰卧位和直立位的六个分区中的两个分区显示出不同的结果(P <.05)。与口腔外相比,Periotest在仰卧位的所有六个部位和直立位的五个部位显示出不同的结果(P <.05)。虽然Anycheck在仰卧位的五个区域与口腔外相比显示出不同的结果,但在直立位仅在一个区域显示出不同的结果(P <.05)。
在手术医生难以触及的种植体位置,种植体稳定性测量装置的可靠性较低。种植体的可达性受Osstell、Anycheck和Periotest的影响程度依次降低。