Lee Dong-Hoon, Shin Yo-Han, Park Jin-Hong, Shim Ji-Suk, Shin Sang-Wan, Lee Jeong-Yol
Department of Prosthodontics, Institute for Clinical Dental Research, Korea University Medicine, Korea University, Seoul, Republic of Korea.
J Adv Prosthodont. 2020 Apr;12(2):83-88. doi: 10.4047/jap.2020.12.2.83. Epub 2020 Apr 22.
The purpose of this study was to examine the reliability of the Anycheck device and the effect of the healing abutment diameter on the Anycheck values (implant stability test, IST).
Thirty implants were placed into three artificial bone blocks with 10 Ncm, 15 Ncm, and 35 Ncm insertion torque value (ITV), respectively (n = 10). (1) The implant stability was measured with three different kinds of devices (Periotest M, Osstell ISQ Mentor, and Anycheck). (2) Five different diameters (4.0, 4.5, 4.8, 5.5, and 6.0 mm) of healing abutments of the same height were connected to the implants and the implant stability was measured four times in different directions with Anycheck. The measured mean values were statistically analyzed.
The correlation coefficient between the mean implant stability quotient (ISQ) and IST value was 0.981 (<.01) and the correlation coefficient between the meant periotest value (PTV) and IST value was -0.931 (<.01). There were no statistically significant differences among the IST values with different healing abutment diameters.
There was a strong correlation between the Periotest M and Anycheck values and between the ISQ and IST. The diameter of the healing abutment had no effect on the Anycheck values.
本研究旨在检验Anycheck设备的可靠性以及愈合基台直径对Anycheck值(种植体稳定性测试,IST)的影响。
将30枚种植体分别以10 Ncm、15 Ncm和35 Ncm的植入扭矩值(ITV)植入三块人工骨块中(n = 10)。(1)使用三种不同的设备(Periotest M、Osstell ISQ Mentor和Anycheck)测量种植体稳定性。(2)将五种不同直径(4.0、4.5、4.8、5.5和6.0 mm)且高度相同的愈合基台连接到种植体上,并用Anycheck在不同方向测量种植体稳定性四次。对测量的平均值进行统计学分析。
种植体稳定性商数(ISQ)平均值与IST值之间的相关系数为0.981(<.01),Periotest值(PTV)平均值与IST值之间的相关系数为 -0.931(<.01)。不同愈合基台直径的IST值之间无统计学显著差异。
Periotest M值与Anycheck值之间以及ISQ与IST之间存在强相关性。愈合基台直径对Anycheck值无影响。