利用新型压缩装置评估颈椎椎间孔变化的 MRI 研究。

MRI evaluation of foraminal changes in the cervical spine with assistance of a novel compression device.

机构信息

Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.

Department of Orthopaedics, Sahlgrenska University Hospital, Gothenburg, Sweden.

出版信息

Sci Rep. 2023 Jul 17;13(1):11508. doi: 10.1038/s41598-023-38401-5.

Abstract

Standard supine Magnetic Resonance Imaging (MRI) does not acquire images in a position where most patients with intermittent arm radiculopathy have symptoms. The aim of this study was to test the feasibility of a new compression device and to evaluate image quality and foraminal properties during a Spurling test under MRI acquisition. Ten asymptomatic individuals were included in the study (6 men and 4 women; age range 27 to 55 years). First, the subjects were positioned in the cervical compression device in a 3 T MRI scanner, and a volume T2 weighted (T2w) sequence was acquired in a relaxed supine position (3 min). Thereafter, the position and compressive forces on the patient's neck (provocation position) were changed by maneuvering the device from the control room, with the aim to simulate a Spurling test, causing a mild foraminal compression, followed by a repeated image acquisition (3 min). A radiologist measured the blinded investigations evaluating cervical lordosis (C3-C7), foraminal area on oblique sagittal images and foraminal cross-distance in the axial plane. A total of three levels (C4-C7) were measured on the right side on each individual. Measurements were compared between the compressed and relaxed state. Reliability tests for inter- and intraclass correlation were performed. The device was feasible to use and well tolerated by all investigated individuals. Images of adequate quality was obtained in all patients. A significant increase (mean 9.4, p = 0.013) in the cervical lordosis and a decreased foraminal cross-distance (mean 32%, p < 0.001) was found, during the simulated Spurling test. The area change on oblique sagittal images did not reach a statistically significant change. The reliability tests on the quantitative measures demonstrated excellent intraobserver reliability and moderate to good interobserver reliability. Applying an individualized provocation test on the cervical spine, which simulates a Spurling test, during MRI acquisition was feasible with the novel device and provided images of satisfactory quality. MRI images acquired with and without compression showed changes in cervical lordosis and foraminal cross distance indicating the possibility of detecting changes of the foraminal properties. As a next step, the method is to be tested on symptomatic patients.

摘要

标准仰卧位磁共振成像(MRI)无法在间歇性臂神经根病患者出现症状的位置采集图像。本研究的目的是测试一种新的压缩设备的可行性,并在 MRI 采集过程中评估 Spurling 试验期间的图像质量和椎间孔特性。研究纳入了 10 名无症状个体(6 名男性和 4 名女性;年龄 27 至 55 岁)。首先,受试者在 3T MRI 扫描仪中使用颈椎压缩设备定位,并在放松的仰卧位采集 T2 加权(T2w)序列(3 分钟)。此后,通过在控制室操作设备来改变患者颈部的位置和压缩力(激发位置),旨在模拟 Spurling 试验,引起轻度椎间孔压缩,然后重复图像采集(3 分钟)。放射科医生对评估颈椎曲度(C3-C7)、斜矢状位图像上的椎间孔面积和轴向平面上的椎间孔横距的盲法研究进行了测量。在每个个体的右侧共测量了三个水平(C4-C7)。比较了压缩状态和放松状态下的测量值。进行了组内和组间相关的可靠性测试。该设备易于使用,所有被调查者均能耐受。所有患者均获得了足够质量的图像。在模拟 Spurling 试验期间,发现颈椎曲度显著增加(平均增加 9.4,p=0.013),椎间孔横距减小(平均 32%,p<0.001)。斜矢状位图像上的面积变化未达到统计学显著变化。定量测量的可靠性测试显示观察者内的可靠性极好,观察者间的可靠性为中度至良好。使用新设备在 MRI 采集过程中对颈椎进行个体化激发试验,模拟 Spurling 试验是可行的,并提供了质量满意的图像。在有和没有压缩的情况下采集的 MRI 图像显示颈椎曲度和椎间孔横距的变化,表明有可能检测到椎间孔特性的变化。下一步是在有症状的患者中测试该方法。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/f80b/10352332/ca5ecf5640b6/41598_2023_38401_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验