Department of Laboratory Medicine, Europe Hospitals, Brussels, Belgium.
Scand J Clin Lab Invest. 2023 Oct;83(6):390-393. doi: 10.1080/00365513.2023.2241356. Epub 2023 Jul 28.
This study investigates the impact of sample type on the measurement of 25-OH-vitamin D using the Liaison XL (Diasorin) and Cobas e801 (Roche). This investigation was motivated by the need to optimize sample volume usage, which led us to adopt the use of heparin plasma, an alternative proposed by Diasorin in their specification. Discordant and unexplainable results were observed, prompting us to evaluate the effect of sample type on the accuracy of the 25-OH-vitamin D measurements. We collected 34 different paired samples from a randomly selected patients who had two types of tubes taken simultaneously: serum-gel and lithium-heparin plasma tubes. The 25-OH-vitamin D levels were measured using Cobas e801 and Liaison. Statistical analysis was performed using the Mann-Whitney test to calculate the p-value. Biases were also calculated. When comparing the heparin matrix with the serum matrix on the Liaison XL analyzer, a higher proportion ( < .0001; 79% versus 64%) of patients were classified in the 'normal group', while fewer were classified in the 'insufficiency' or 'deficiency group'. The heparin tubes on the Liaison XL analyzer showed a mean bias of 57.5%) (-value < .001; 95%CI: 37.6-77.4) compared to the serum tubes. On the other hand, the heparin tubes on the Cobas e801 analyzer showed a mean bias of -0.2% (95%CI: -4.8 to 4.5) compared to the serum tubes. It is imperative for laboratory professionals to be aware of this interference for an accurate measurement of 25-OH-vitamin D levels on the Liaison XL. Further research is needed to understand the mechanism of this interference.
本研究旨在探讨使用 Liaison XL(Diasorin)和 Cobas e801(Roche)检测 25-OH-维生素 D 时样本类型的影响。我们之所以进行此项研究,是因为需要优化样本量的使用,这促使我们采用肝素血浆作为替代样本,这是 Diasorin 在其说明书中提出的一种选择。我们观察到了不一致且无法解释的结果,这促使我们评估样本类型对 25-OH-维生素 D 测量准确性的影响。我们从随机选择的患者中收集了 34 对不同的样本,这些患者同时采集了两种类型的管:血清凝胶管和锂肝素血浆管。使用 Cobas e801 和 Liaison 测量 25-OH-维生素 D 水平。使用 Mann-Whitney 检验进行统计分析,计算 p 值。还计算了偏差。当在 Liaison XL 分析仪上比较肝素基质与血清基质时,较高比例(<0.0001;79%比 64%)的患者被归类为“正常组”,而较少的患者被归类为“不足”或“缺乏组”。与血清管相比,肝素管在 Liaison XL 分析仪上的平均偏差为 57.5%(-值<0.001;95%CI:37.6-77.4)。另一方面,与血清管相比,肝素管在 Cobas e801 分析仪上的平均偏差为-0.2%(95%CI:-4.8 至 4.5)。实验室专业人员必须意识到这种干扰,以准确测量 Liaison XL 上的 25-OH-维生素 D 水平。需要进一步研究以了解这种干扰的机制。