Turkes A, Turkes A O, Joyce B G, Read G F, Riad-Fahmy D
Steroids. 1979 Mar;33(3):347-59. doi: 10.1016/0039-128x(79)90011-4.
A sensitive, solid phase enzymeimmunoassay suitable for determining testosterone concentrations in samll aliquots of plasma (20 microliter) and saliva (200 microliter) has been developed. A solid phase antiserum raised against a testosterone-11 alpha-hemisuccinate/bovine serum albumin conjugate was prepared by coupling to cyanogen bromide activated cellulose. The "enzyme label" was a covalently linked testosterone/horseradish peroxidase conjugate. The assay had a lower limit of sensitivity of 4pg/assay tube and satisfied accepted criteria of specificity and precision. Testosterone concentrations determined by enzyme-immunoassay were in excellent agreement not only with a gas liquid chromatography/mass spectrometry procedure (r=0.96, n=12) but also with the radioimmunoassay in routine use (r=0.95, n=12). The EIA can therefore replace RIA in both the small clinical laboratory and high throughput service centres for determining plasma and salivary testosterone concentrations. In normal males salivary testosterone concentrations reflected circulating steroid levels and indicated the possibility of assaying saliva rather than plasma in clinical studies.
已开发出一种灵敏的固相酶免疫测定法,适用于测定少量血浆(20微升)和唾液(200微升)中的睾酮浓度。通过与溴化氰活化纤维素偶联,制备了针对睾酮-11α-半琥珀酸酯/牛血清白蛋白缀合物的固相抗血清。“酶标记物”是一种共价连接的睾酮/辣根过氧化物酶缀合物。该测定法的最低检测限为4皮克/测定管,满足特异性和精密度的公认标准。通过酶免疫测定法测定的睾酮浓度不仅与气相色谱/质谱法(r = 0.96,n = 12)高度一致,而且与常规使用的放射免疫测定法(r = 0.95,n = 12)高度一致。因此,在小型临床实验室和高通量服务中心,酶免疫测定法均可替代放射免疫测定法来测定血浆和唾液中的睾酮浓度。在正常男性中,唾液睾酮浓度反映了循环类固醇水平,并表明在临床研究中检测唾液而非血浆的可能性。