Effects of Cuff Size on the Accuracy of Blood Pressure Readings: The Cuff(SZ) Randomized Crossover Trial.

机构信息

Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland.

Johns Hopkins University School of Medicine, Baltimore, Maryland.

出版信息

JAMA Intern Med. 2023 Oct 1;183(10):1061-1068. doi: 10.1001/jamainternmed.2023.3264.

Abstract

IMPORTANCE

Clinical practice guidelines recommend selecting an appropriately sized cuff based on mid-arm circumference prior to measuring blood pressure (BP). To our knowledge, the effect of miscuffing on BP measurement when using an automated BP device has not been quantified.

OBJECTIVE

To determine the effect of using a regular BP cuff vs an appropriately sized BP cuff on automated BP readings.

DESIGN, SETTING, AND PARTICIPANTS: This randomized crossover trial of community-dwelling adults with a wide range of mid-arm circumferences took place between March 16 and October 25, 2021, in Baltimore, Maryland. Participants were recruited via BP screening events at a public food market and a senior housing facility, targeted mailings to prior research participants, placement of study brochures in hypertension clinics at Johns Hopkins University, and referrals from physicians providing hypertension care to adults.

INTERVENTIONS

Participants underwent 4 sets of triplicate BP measurements, with the initial 3 sets using an appropriate, too-small, or too-large BP cuff in random order; the fourth set of triplicate measurements was always completed with an appropriate BP cuff.

MAIN OUTCOMES AND MEASURES

The primary outcome was the difference in mean BP when measured with a regular BP cuff compared with an appropriate BP cuff. The secondary outcome was the difference in BP when using too-small or too-large BP cuffs vs an appropriate BP cuff across all cuff sizes. Results were also stratified by systolic BP (≥130 mm Hg vs <130 mm Hg) and body mass index (calculated as weight in kilograms divided by height in meters squared; ≥30 vs <30).

RESULTS

A total of 195 adults (mean [SD] age, 54 [16] years; 67 [34%] male; 132 [68%] Black; 100 [51%] with hypertension) were randomized for inclusion. Among individuals requiring a small BP cuff, use of a regular BP cuff resulted in a statistically significant lower BP reading (mean systolic BP difference, -3.6 [95% CI, -5.6 to -1.7] mm Hg). In contrast, among individuals requiring a large or extra-large BP cuff, use of a regular BP cuff resulted in a statistically significant higher BP reading (mean systolic BP difference, 4.8 [95% CI, 3.0-6.6] mm Hg and 19.5 [95% CI, 16.1-22.9] mm Hg, respectively). For the secondary outcome, BP differences with overcuffing and undercuffing by 1 and 2 cuff sizes were greater among those requiring larger BP cuffs. The results were consistent in stratified analyses by systolic BP and body mass index.

CONCLUSIONS AND RELEVANCE

In this randomized crossover trial, miscuffing resulted in strikingly inaccurate BP measurements. This is particularly concerning for settings where 1 regular BP cuff size is routinely used in all individuals, regardless of arm size. A renewed emphasis on individualized BP cuff selection is warranted.

TRIAL REGISTRATION

ClinicalTrials.gov Identifier: NCT04610775.

摘要

重要性

临床实践指南建议在测量血压(BP)之前,根据中臂周长选择合适尺寸的袖带。据我们所知,使用自动 BP 设备时,袖带不合适对 BP 测量的影响尚未量化。

目的

确定使用常规 BP 袖带与合适尺寸的 BP 袖带对自动 BP 读数的影响。

设计、地点和参与者:这项针对中臂周长范围广泛的社区居住成年人的随机交叉试验于 2021 年 3 月 16 日至 10 月 25 日在马里兰州巴尔的摩进行。参与者通过公共食品市场和高级住房设施的 BP 筛查活动、针对先前研究参与者的定向邮件、在约翰霍普金斯大学高血压诊所放置研究小册子以及提供成人高血压护理的医生转介招募。

干预措施

参与者接受了 4 组 3 次重复的 BP 测量,前 3 组以随机顺序分别使用合适、太小或太大的 BP 袖带;第 4 组 3 次重复测量始终使用合适的 BP 袖带完成。

主要结果和措施

主要结果是使用常规 BP 袖带与合适 BP 袖带测量时平均 BP 的差异。次要结果是在所有袖带尺寸下,使用太小或太大的 BP 袖带与合适的 BP 袖带相比 BP 的差异。结果还按收缩压(≥130mmHg 与<130mmHg)和身体质量指数(体重以千克为单位除以身高以米为单位的平方;≥30 与<30)进行分层。

结果

共有 195 名成年人(平均[标准差]年龄,54[16]岁;67[34%]男性;132[68%]黑人;100[51%]患有高血压)被随机纳入。在需要小 BP 袖带的人群中,使用常规 BP 袖带导致血压读数明显降低(平均收缩压差异,-3.6[95%CI,-5.6 至-1.7]mmHg)。相比之下,在需要大或特大 BP 袖带的人群中,使用常规 BP 袖带导致血压读数明显升高(平均收缩压差异分别为 4.8[95%CI,3.0-6.6]mmHg 和 19.5[95%CI,16.1-22.9]mmHg)。对于次要结果,袖带尺寸过大或过小 1 或 2 个尺寸的 BP 差异在需要更大 BP 袖带的人群中更大。分层分析中按收缩压和身体质量指数进行分析,结果一致。

结论和相关性

在这项随机交叉试验中,袖带不合适导致 BP 测量结果明显不准确。对于所有个体无论手臂大小都常规使用 1 个常规 BP 袖带尺寸的情况,这尤其令人担忧。有必要重新强调个体化 BP 袖带选择。

试验注册

ClinicalTrials.gov 标识符:NCT04610775。

相似文献

引用本文的文献

6
[Not Available].[无可用内容]
CMAJ. 2025 Jun 15;197(23):E656-E673. doi: 10.1503/cmaj.241770-f.
9
Invasive and non-invasive monitoring in the ICU.重症监护病房中的有创和无创监测
Trauma Surg Acute Care Open. 2025 Apr 14;10(Suppl 1):e001780. doi: 10.1136/tsaco-2025-001780. eCollection 2025.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验