Ribeiro Débora de Fátima Camillo, Hembecker Paula Karina, Nakato Adriane Muller, Fernandes Beatriz Luci, Nohama Percy
Neonatal Services, Waldemar Monastier Hospital, Campo Largo, Paraná, Brazil.
Graduate Program on Health Technology, Pontifícia Universidade Católica Do Paraná, Curitiba, Paraná, Brazil.
J Multidiscip Healthc. 2023 Aug 2;16:2217-2229. doi: 10.2147/JMDH.S415748. eCollection 2023.
To evaluate the health professionals' satisfaction regarding the short binasal prong used in Neonatal Intensive Care Units (NICU), evaluate the difficulties related to its use, and present possible improvements in the design of this device.
Observational, cross-sectional study with prospective data collection carried out in the NICU of a public hospital in southern Brazil. This research was presented into two stages. In the first stage, the Quebec User Evaluation of Satisfaction with Assistive Technology was applied with 90 health professionals to evaluate the satisfaction regarding the short binasal prong. In the second stage, the health professional's experiences and difficulties in using the binasal prongs in 14 newborns (NBs) that required non-invasive ventilation was collected. The short binasal prongs used was the Fanem brand and the CPAP circuit was Gabisa Medical International (GMI). Finally, improvements and recommendations were presented to optimize the device's design.
The mean score of the health professionals' satisfaction with short binasal prongs was 3.8 ± 0.6. Ease of adjustment (3.27) and dimensions (3.62) variables had the worst scores. The main difficulties pointed out by health professionals were: circuit disconnection (57.1%), the size of the prong did not correspond to NBs' anatomical characteristics (35.7%), air leakage (21.4%), and difficulty in fixing and positioning the prong in the NB (14.28%). The improvements suggested were: appropriate prong sizes based on the anatomical characteristics of the NBs; adjustable distance between insertion and base catheters; manufactured with malleable material, however not easily foldable; curved and adjustable insertion catheters and functional system of tracheas' connection.
The dissatisfaction of health professionals with the dimensions and prongs adjustments and the difficulties faced in clinical practice indicate the need for improvements in these interfaces. The recommendations presented in this study may contribute to optimizing the design of the binasal prong in the future.