Hosoba M, Wani H, Hiroe M, Kusakabe K
Eur J Nucl Med. 1986;12(2):53-9. doi: 10.1007/BF00364728.
A new fully-automated method for processing gated blood-pool images is presented and its clinical validity and performance for images with various noise levels are investigated using data obtained from a slant-hole collimator. The optimal preprocessing conditions are evaluated for the images with various noise levels to obtain the accurate ejection fraction, end-diastolic and end-systolic counts. This new method has successfully detected left ventricular contour in 92% of the 61 patients. The left ventricular ejection fraction obtained by the method related closely to that of contrast ventriculography (correlation coefficient r = 0.90). The end-diastolic volume also had a good correlation with contrast ventriculography (r = 0.90).
本文介绍了一种用于处理门控血池图像的全新全自动方法,并使用从斜孔准直器获得的数据,研究了其在不同噪声水平图像上的临床有效性和性能。针对不同噪声水平的图像评估了最佳预处理条件,以获得准确的射血分数、舒张末期和收缩末期计数。该新方法在61例患者中的92%成功检测到了左心室轮廓。通过该方法获得的左心室射血分数与对比心室造影密切相关(相关系数r = 0.90)。舒张末期容积与对比心室造影也具有良好的相关性(r = 0.90)。