Suppr超能文献

一种用于分析冠状动脉管径和血流的微型计算机系统。

A microcomputer system for analysis of coronary artery calibre and blood flow.

作者信息

Sheridan D J, Culling W, Thomas P, Flores N

出版信息

J Med Eng Technol. 1986 May-Jun;10(3):121-5. doi: 10.3109/03091908609022898.

Abstract

Measurement of myocardial blood flow and assessment of coronary calibre in man has important clinical and research value. An inexpensive microcomputer system has been developed to facilitate analysis of each of these. Coronary sinus and great cardiac vein blood flow are measured using the thermodilution technique. The output of the temperature measurement circuitry is digitized and used to calculate blood flows. For coronary artery calibre measurement, the outline of an arterial segment recorded during coronary arteriography is digitized manually. The microcomputer system provides a graphic display of the digitized artery and calculates diameters throughout its length. Excellent correlations were observed between computer and manual methods for both systems.

摘要

测量人体心肌血流量和评估冠状动脉管径具有重要的临床和研究价值。已开发出一种价格低廉的微型计算机系统,以方便对上述各项进行分析。采用热稀释技术测量冠状窦和心大静脉血流量。温度测量电路的输出被数字化,并用于计算血流量。对于冠状动脉管径测量,冠状动脉造影期间记录的动脉节段轮廓通过手动数字化。微型计算机系统提供数字化动脉的图形显示,并计算其全长的直径。两种系统的计算机方法和手动方法之间均观察到极好的相关性。