Arch Soc Esp Oftalmol (Engl Ed). 2023 Sep;98(9):533-539. doi: 10.1016/j.oftale.2023.07.001. Epub 2023 Aug 16.
This study analyzes the comparability of measurements taken by a Retinomax K-plus 3 handheld autorefractometer in Quick mode and a Topcon KR-800 on-table autorefractometer in standard mode on the pediatric population, and establishes their correlation.
It is a retrospective comparative study. Spherical diopter power (SPH), cylindrical diopter power (CYL), angle of cylindrical axis (AX), and spherical equivalent (SE) were measured with the Retinomax in Quick mode and with the Topcon in standard mode. Each patient was evaluated in cycloplegic and non-cycloplegic conditions by both autorefractometers. Student's t-test was performed between the two instruments for SPH, CYL, and SE. The Pearson correlation coefficient was calculated and the dispersion was represented using Bland-Altman graphs, also evaluating the subgroup of patients under 4 years of age. A descriptive analysis of the percentages of measures that differed was performed.
It included 98 eyes of 49 subjects (age range: 3-16 years). The data for HPS without cycloplegia are virtually identical, whereas with cycloplegia there is a hyperopic bias of +0.5 diopters measured with Retinomax. CYL results are very similar with and without cycloplegia. There is a high Pearson correlation for both instruments (>0.91) and a low degree of dispersion in the Bland-Altman plots under cycloplegia.
The Retinomax data were consistent with those obtained by Topcon. The Retinomax is a useful instrument for detecting refractive errors in children between 3 and 16 years of age.
本研究分析了 Retinomax K-plus 3 手持式自动验光仪 Quick 模式和 Topcon KR-800 桌上式自动验光仪标准模式在儿科人群中的测量值的可比性,并建立了它们的相关性。
这是一项回顾性比较研究。使用 Retinomax 的 Quick 模式和 Topcon 的标准模式测量球镜度数(SPH)、柱镜度数(CYL)、柱镜轴角度(AX)和等效球镜度(SE)。每个患者均在睫状肌麻痹和非睫状肌麻痹状态下由两种自动验光仪进行评估。对 Retinomax 和 Topcon 的 SPH、CYL 和 SE 进行了 Student's t 检验。计算了 Pearson 相关系数,并通过 Bland-Altman 图表示了离散度,还评估了 4 岁以下患者的亚组。对差异测量值的百分比进行了描述性分析。
纳入 49 名患者(年龄范围:3-16 岁)的 98 只眼。未散瞳时 HPS 的数据几乎相同,而 Retinomax 测量时则有+0.5 屈光度的远视偏差。散瞳和非散瞳时 CYL 结果非常相似。两种仪器的 Pearson 相关系数均>0.91,且散瞳时 Bland-Altman 图的离散度较低。
Retinomax 的数据与 Topcon 一致。Retinomax 是一种用于检测 3-16 岁儿童屈光不正的有用仪器。