Abstract

BACKGROUND

Carbohydrate antigen 724 (CA724) is a sensitive and specific indicator for multiple malignant tumors. The aim of this study was to establish a Eu-time resolved fluorescence immunochromatography (Eu-TRFICO) method for quantitative detection of CA724 in serum.

METHODS

Eu-TRFICO strips were optimized and assembled. The sensitivity, specificity and precision were evaluated using CA724 standard dilutions and matrix serum. Meanwhile, the reference interval, comparison, and sensitivity/specificity were performed using clinical negative/positive gastric cancer serum samples.

RESULTS

The standard curve equation was y = 9.869 x - 154.12 (R = 0.993), and the sensitivity was 0.42 U/mL. The common interferents in serum could not affect the quantitative results with low cross-reactivities (all no more than 1.09%). All average recoveries of the intra- and interbatch ranged from 102.38 to 106.40%, and all CVs were below 10%. The reference interval of the healthy subjects was < 4.68 U/mL and the reference interval of the subjects with grade I/II gastric cancer was > 9.54 U/mL. Additionally, a high Pearson r (0.9503) and sensitivity/specificity (92.86%/94.20%) were obtained.

CONCLUSION

This study prepared Eu-TRFICO strips with high sensitivity, specificity, precision and satisfactory clinical testing performance, which provides more options for clinical quantitative and convenient testing of CA724.

摘要

背景

糖类抗原 724(CA724)是多种恶性肿瘤的敏感且特异的指标。本研究旨在建立一种用于血清中 CA724 定量检测的 Eu-时间分辨荧光免疫层析(Eu-TRFICO)方法。

方法

优化并组装 Eu-TRFICO 条。使用 CA724 标准稀释液和基质血清评估灵敏度、特异性和精密度。同时,使用临床阴性/阳性胃癌血清样本进行参考区间、比较和灵敏度/特异性分析。

结果

标准曲线方程为 y=9.869x-154.12(R=0.993),灵敏度为 0.42 U/mL。血清中的常见干扰物不会影响定量结果,交叉反应率均较低(均不超过 1.09%)。批内和批间的所有平均回收率均在 102.38%至 106.40%之间,所有 CV 均低于 10%。健康受试者的参考区间<4.68 U/mL,I/II 级胃癌受试者的参考区间>9.54 U/mL。此外,还获得了高度的 Pearson r(0.9503)和灵敏度/特异性(92.86%/94.20%)。

结论

本研究制备了具有高灵敏度、特异性、精密度和令人满意的临床检测性能的 Eu-TRFICO 条,为 CA724 的临床定量和便捷检测提供了更多选择。