Hui Terrence C H, How Guo Yuan, Chim Michelle S M, Pua Uei
Department of Diagnostic Radiology, Tan Tock Seng Hospital, Singapore 308433, Singapore.
Yong Loo Lin School of Medicine, National University of Singapore, Singapore 117597, Singapore.
Diagnostics (Basel). 2023 Aug 18;13(16):2702. doi: 10.3390/diagnostics13162702.
(1) Background: Percutaneous microwave ablation (MWA) is an accepted treatment of non-operative liver cancer. This study compares the ablation zones of four commercially available 2.45 GHz MWA systems (Emprint, Eco, Neuwave, and Solero) in an ex vivo porcine liver model. (2) Methods: Ex vivo porcine livers ( = 85) were obtained. Two ablation time setting protocols were evaluated, the manufacturer's recommended maximum time and a 3 min time, performed at the manufacturer-recommended maximum power setting. A total of 236 ablation samples were created with 32 (13.6%) samples rejected. A total of 204 samples were included in the statistical analysis. (3) Results: For single-probe protocols, Emprint achieved ablation zones with the largest SAD. Significant differences were found in all comparisons for the 3 min time setting and for all comparisons at the 10 min time setting except versus Neuwave LK15 and Eco. Emprint produced ablation zones that were also significantly more spherical (highest SI) than the single-probe ablations from all other manufacturers. No statistical differences were found for ablation shape or SAD between the single-probe protocols for Emprint and the three-probe protocols for Neuwave. (4) Conclusions: The new Emprint HP system achieved large and spherical ablation zones relative to other 2.45 GHz MWA systems.
(1) 背景:经皮微波消融(MWA)是一种被认可的非手术治疗肝癌的方法。本研究在离体猪肝模型中比较了四种市售2.45GHz MWA系统(Emprint、Eco、Neuwave和Solero)的消融区域。(2) 方法:获取离体猪肝(n = 85)。评估了两种消融时间设置方案,即制造商推荐的最长时间和3分钟时间,均在制造商推荐的最大功率设置下进行。共创建了236个消融样本,其中32个(13.6%)样本被排除。共有204个样本纳入统计分析。(3) 结果:对于单探针方案,Emprint实现的消融区域具有最大的短轴直径(SAD)。在3分钟时间设置的所有比较以及10分钟时间设置的所有比较中(除与Neuwave LK15和Eco比较外)均发现显著差异。Emprint产生的消融区域也比所有其他制造商的单探针消融区域更接近球形(球形指数最高)。Emprint的单探针方案与Neuwave的三探针方案在消融形状或SAD方面未发现统计学差异。(4) 结论:相对于其他2.45GHz MWA系统,新型Emprint HP系统实现了大且呈球形的消融区域。