A comparison of three different microwave systems for laparoscopic liver tumor ablation.

机构信息

Department of Endocrine Surgery, Cleveland Clinic, Cleveland, Ohio, USA.

Department of General Surgery, Cleveland Clinic, Cleveland, Ohio, USA.

出版信息

J Surg Oncol. 2024 Jun;129(7):1245-1253. doi: 10.1002/jso.27621. Epub 2024 Mar 12.

Abstract

BACKGROUND

Our aim was to perform a comparison of three current microwave ablation (MWA) systems widely used for laparoscopic liver ablations in terms of ablation kinetics and geometry of ablation zones.

METHODS

This was a retrospective, institutional review board-approved study comparing Emprint, Emprint HP, and NeuWave systems for laparoscopic liver ablation. Analyses were performed via Mann-Whitney U and χ tests. Continuous data are presented as median (interquartile range).

RESULTS

For Emprint, Emprint HP, and NeuWave groups, tumor size was 1.16 (0.8), 1.21 (0.7), and 1.27 (0.9) cm (p = 0.54). Ablation time per lesion was 7 (6), 4 (2.8), and 4 (3.3) min (p < 0.0001), yielding similar ablation zone volumes and margins. The time to first ablation bubble was 1 (0.13), 1.5 (0.85), and 0.75 (0.5) min, and total ablation times were 7 (4.4), 4 (2), and 3.5 (2.8) min (p < 0.0001). The roundness index A, B, and transverse were 0.94, 0.98, and 0.79; 0.95, 0.95, and 0.78; and 1.02, 0.95, and 0.96.

CONCLUSIONS

Although a saline-cooling system with Emprint system allowed for larger diameter spherical ablation zones to be created, it led to decreased efficiency compared to the CO-cooled NeuWave system, which exposes the active antenna directly to tissue. Increased power delivered by Emprint HP improved the efficiency of saline-cooled design, as demonstrated by faster ablation times.

摘要

背景

本研究旨在比较三种广泛应用于腹腔镜肝脏消融的当前微波消融(MWA)系统,以评估消融动力学和消融区域的几何形状。

方法

这是一项回顾性的、经机构审查委员会批准的研究,比较了 Emprint、Emprint HP 和 NeuWave 系统用于腹腔镜肝脏消融的效果。分析采用 Mann-Whitney U 和 χ 检验。连续数据以中位数(四分位间距)表示。

结果

Emprint、Emprint HP 和 NeuWave 组的肿瘤大小分别为 1.16(0.8)、1.21(0.7)和 1.27(0.9)cm(p=0.54)。每个病灶的消融时间分别为 7(6)、4(2.8)和 4(3.3)min(p<0.0001),产生了相似的消融区域体积和边界。首次消融气泡的时间分别为 1(0.13)、1.5(0.85)和 0.75(0.5)min,总消融时间分别为 7(4.4)、4(2)和 3.5(2.8)min(p<0.0001)。圆形指数 A、B 和横径分别为 0.94、0.98 和 0.79;0.95、0.95 和 0.78;1.02、0.95 和 0.96。

结论

尽管 Emprint 系统的盐水冷却系统允许创建更大直径的球形消融区域,但与将活性天线直接暴露于组织的 CO 冷却 NeuWave 系统相比,其效率降低。Emprint HP 增加的功率提高了盐水冷却设计的效率,因为消融时间更快。