It has been reported that cerebral oxygen saturation (rSO) measured by near infrared spectroscopy is low in dialysis patients. We compared the rSO values of dialysis patients before living donor kidney transplantation and their donors as controls by using three spectroscopes that utilize different principals, the INVOS 5100C (spatially resolved spectroscopy), FORE-SIGHT ELITE (modified Beer-Lambert law) and tNIRS-1 (time-resolved spectroscopy).
Before induction of anesthesia, the sensors of one of the three spectroscopes were placed on the forehead and rSO values were recorded followed by the same measurement using the other two spectroscopes. The primary objective was to compare the rSO values of the dialysis patients and controls using the three spectroscopes by the unpaired t test. Then we compared the rSO values among the spectroscopes in both dialysis patients and controls by one-way ANOVA. Finally, we examined the relations between the rSO values and the physiological values by using the Pearson correlation coefficient.
Fifteen pairs of dialysis patients and controls were studied. With the INVOS 5100 C, the values of the dialysis patients (59.7 ± 9.7% (mean ± standard deviation) were 13% lower than those of the controls (73.3 ± 6.9%) (P < 0.01). With the tNIRS-1, the values were 57.8 ± 4.8% in the dialysis patients and 63.3 ± 3.5% in the controls (P < 0.01). Almost no differences were observed with the FORE-SIGHT ELITE (71.6 ± 4.9% [dialysis patients] vs. 70.8 ± 4.3% [Controls]) (P = 0.62). Among the spectroscopes, the values were significantly different in both dialysis patients and controls. For the INVOS 5100C and tNIRS-1, correlation coefficients between rSO values and blood Hb and serum Alb were more than 0.5.
The rSO values for comparisons between the dialysis patients and the controls were different according to differences of the principles of the near infrared spectroscopes. In the INVOS 5100C and tNIRS-1, rSO values may be related to blood Hb and serum Alb.
据报道,近红外光谱仪测量的脑氧饱和度(rSO)在透析患者中较低。我们通过使用三种利用不同原理的光谱仪(空间分辨光谱仪 INVOS 5100C、改良 Beer-Lambert 定律的 FORE-SIGHT ELITE 和时分辨光谱仪 tNIRS-1),比较了接受活体供肾移植前透析患者及其供者的 rSO 值。
在麻醉诱导前,将三种光谱仪中的一种的传感器放置在前额上,记录 rSO 值,然后使用另外两种光谱仪进行相同的测量。主要目的是通过配对 t 检验比较透析患者和对照组使用三种光谱仪的 rSO 值。然后,我们通过单向方差分析比较了透析患者和对照组中三种光谱仪的 rSO 值。最后,我们使用 Pearson 相关系数检查 rSO 值与生理值之间的关系。
研究了 15 对透析患者和对照组。使用 INVOS 5100C,透析患者的 rSO 值(59.7±9.7%(均值±标准差))比对照组(73.3±6.9%)低 13%(P<0.01)。使用 tNIRS-1,透析患者的 rSO 值为 57.8±4.8%,对照组为 63.3±3.5%(P<0.01)。使用 FORE-SIGHT ELITE 几乎没有差异(透析患者 71.6±4.9%[vs.对照组 70.8±4.3%])(P=0.62)。在三种光谱仪中,透析患者和对照组的 rSO 值均有显著差异。对于 INVOS 5100C 和 tNIRS-1,rSO 值与血液 Hb 和血清 Alb 之间的相关系数大于 0.5。
根据近红外光谱仪原理的差异,透析患者与对照组之间的 rSO 值比较结果不同。在 INVOS 5100C 和 tNIRS-1 中,rSO 值可能与血液 Hb 和血清 Alb 有关。