J Neurosurg Anesthesiol. 2003 Oct;15(4):302-6. doi: 10.1097/00008506-200310000-00002.
Cerebral oximeter based on near-infrared spectroscopy has been used as a continuous, noninvasive monitoring of regional cerebral oxygen saturation (rSO2). Although the absolute rSO2 values have a wide range of variability, the factors affecting a variability of rSO2 values have not been extensively investigated. The authors investigated the influence of patient variables and sensor location on rSO2 measured by the cerebral oximeter INVOS 4100 in 111 patients anesthetized with sevoflurane, fentanyl, and nitrous oxide in oxygen. The sensors for rSO2 measurements were applied on the right forehead (R), 1 cm lateral to R (R1), on the left forehead (L), 1 cm lateral to L (L1), and on the center of the forehead (C). The relationship between the rSO2 values and patient variables were also analyzed. Values of rSO2 at R1 and L1 were significantly lower than those at R and L, respectively. Values of rSO2 at C were significantly higher compared with those at other sites. There were no significant correlations between the rSO2 values and values of weight, height, and head size. Values of rSO2 were similar between males and females. A significant negative correlation between the rSO2 values and age and a positive correlation between the rSO2 values and hemoglobin concentration were observed. These data indicate that patient age, hemoglobin concentration at the measurement, and sensor location can affect rSO2 values.
基于近红外光谱技术的脑血氧饱和度仪已被用于对局部脑氧饱和度(rSO2)进行连续、无创监测。尽管rSO2的绝对值存在很大的变异性,但影响rSO2值变异性的因素尚未得到广泛研究。作者调查了患者变量和传感器位置对111例接受七氟醚、芬太尼和氧化亚氮-氧气麻醉患者使用INVOS 4100脑血氧饱和度仪测量rSO2的影响。用于测量rSO2的传感器分别置于右前额(R)、R外侧1 cm处(R1)、左前额(L)、L外侧1 cm处(L1)以及前额中央(C)。还分析了rSO2值与患者变量之间的关系。R1和L1处的rSO2值分别显著低于R和L处的rSO2值。C处的rSO2值显著高于其他部位。rSO2值与体重、身高和头围之间无显著相关性。男性和女性的rSO2值相似。观察到rSO2值与年龄呈显著负相关,与血红蛋白浓度呈正相关。这些数据表明,患者年龄、测量时的血红蛋白浓度以及传感器位置会影响rSO2值。