0.55T 常规脑部成像的质量评估:临床工作流程中的初步经验。

Quality assessment of routine brain imaging at 0.55 T: initial experience in a clinical workflow.

机构信息

Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.

Department of Biostatistics, University of Michigan, Ann Arbor, Michigan, USA.

出版信息

NMR Biomed. 2024 Jul;37(7):e5017. doi: 10.1002/nbm.5017. Epub 2023 Aug 31.

Abstract

The purpose of this study was to assess the quality of clinical brain imaging in healthy subjects and patients on an FDA-approved commercial 0.55 T MRI scanner, and to provide information about the feasibility of using this scanner in a clinical workflow. In this IRB-approved study, brain examinations on the scanner were prospectively performed in 10 healthy subjects (February-April 2022) and retrospectively derived from 44 patients (February-July 2022). Images collected using the following pulse sequences were available for assessment: axial DWI (diffusion-weighted imaging), apparent diffusion coefficient maps, 2D axial fluid-attenuated inversion recovery images, axial susceptibility-weighted images (both magnitude and phase), sagittal T-weighted (T1w) Sampling Perfection with Application Optimized Contrast images, sagittal T1w MPRAGE (magnetization prepared rapid gradient echo) with contrast enhancement, axial T1w turbo spin echo (TSE) with and without contrast enhancement, and axial T-weighted TSE. Two readers retrospectively and independently evaluated image quality and specific anatomical features in a blinded fashion on a four-point Likert scale, with a score of 1 being unacceptable and 4 being excellent, and determined the ability to answer the clinical question in patients. For each category of image sequences, the mean, standard deviation, and percentage of unacceptable quality images (<2) were calculated. Acceptable (rating ≥ 2) image quality was achieved at 0.55 T in all sequences for patients and 85% of the sequences for healthy subjects. Radiologists were able to answer the clinical question in all patients scanned. In total, 50% of the sequences used in patients and about 60% of the sequences used in healthy subjects exhibited good (rating ≥ 3) image quality. Based on these findings, we conclude that diagnostic quality clinical brain images can be successfully collected on this commercial 0.55 T scanner, indicating that the routine brain imaging protocol may be deployed on this system in the clinical workflow.

摘要

本研究旨在评估在经 FDA 批准的商业 0.55T MRI 扫描仪上对健康受试者和患者进行临床脑部成像的质量,并提供有关在临床工作流程中使用该扫描仪的可行性信息。在这项经过机构审查委员会批准的研究中,对扫描仪上的脑部检查在 2022 年 2 月至 4 月期间前瞻性地进行了 10 名健康受试者,并且在 2022 年 2 月至 7 月期间回顾性地来自 44 名患者。可用于评估的采集到的图像包括:轴位弥散加权成像(diffusion-weighted imaging,DWI)、表观弥散系数图、2 维轴位液体衰减反转恢复图像、轴位磁化率加权图像(幅度和相位)、矢状位 T1 加权(T1w)采样完美应用对比成像、矢状位 T1w MPRAGE(磁化准备快速梯度回波)增强对比度、轴位 T1w 涡轮自旋回波(TSE)增强和不增强对比度,以及轴位 T1w TSE。两名读者以盲法对图像质量和特定解剖特征进行回顾性和独立评估,采用 4 分李克特量表进行评分,1 分为不可接受,4 分为优秀,并确定在患者中回答临床问题的能力。对于每个图像序列类别,计算平均值、标准差和不可接受质量图像(<2)的百分比。在所有患者的所有序列中以及健康受试者 85%的序列中,均可获得可接受(评分≥2)的图像质量。在所有接受扫描的患者中,放射科医生均能够回答临床问题。总的来说,在患者中使用的 50%序列和约 60%的健康受试者中使用的序列表现出良好(评分≥3)的图像质量。基于这些发现,我们得出结论,可在该商用 0.55T 扫描仪上成功采集到诊断质量的临床脑部图像,表明常规脑部成像方案可在该系统中部署在临床工作流程中。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验