即时超声用于监测新生儿脐静脉置管期间导管尖端位置:一项前瞻性研究。

Point-of-care ultrasound for monitoring catheter tip location during umbilical vein catheterization in neonates: a prospective study.

作者信息

Xie Hao-Qiang, Xie Cai-Xuan, Liao Jin-Feng, Xu Feng-Dan, Du Bang, Zhong Bai-Mao, He Xiao-Guang, Li Ning

机构信息

Department of Neonatology, Dongguan Children's Hospital Affiliated to Guangdong Medical University, Dongguan, China.

出版信息

Front Pediatr. 2023 Aug 23;11:1225087. doi: 10.3389/fped.2023.1225087. eCollection 2023.

Abstract

BACKGROUND

Point-of-care ultrasound (POCUS) can guide umbilical vein catheter placement in real time and monitor catheter tip position, allowing avoidance of severe complications due to catheter malposition. This study aims to explore the effectiveness of POCUS in guiding venous catheter insertion and monitoring complications.

METHODS

Sixty-eight neonates with ultrasound-guided venous catheter insertion at the Neonatal Department of Dongguan Children's Hospital between December 2020 and February 2022 were included. POCUS was applied to monitor catheter tip location daily until catheter removal. A displacement range exceeding the intersection of the inferior vena cava and right atrium by ±0.5 cm was considered misalignment.

RESULTS

Sixty-four neonates had a displaced catheter tip (94.1%, 64/68), with a median displacement distance of 0.4 cm (minimum -0.2 cm, maximum 1.2 cm). Ten neonates had a misalignment (14.7%, 10/68) caused by displacement. Displacement usually occurs within 2-4 days after placement, with displacement rates of 94.1% (64/68), 90.6% (58/64), and 98.3% (59/60) on days 2, 3, and 4, respectively, and could still occur on day 9 post-placement. In addition, misalignment mainly occurs on the second day after placement. During the monitoring process, 58 neonates had catheter tip displacement ≥2 times, resulting in 252 displacement and 22 misalignment incidents. Among them, the catheter tip migrated outward from the inferior vena cava seven times, all of which were removed in time. Ultrasound was used for positioning 486 times, and x-ray was indirectly avoided 486 times.

CONCLUSION

The catheter tip is prone to displacement and misalignment after umbilical vein catheterization, which most commonly occurs on days 2-4. POCUS is recommended for daily monitoring of the tip location during umbilical vein catheterization until catheter removal.

摘要

背景

床旁超声(POCUS)可实时引导脐静脉导管置入并监测导管尖端位置,避免因导管位置不当导致的严重并发症。本研究旨在探讨POCUS在引导静脉导管插入及监测并发症方面的有效性。

方法

纳入2020年12月至2022年2月在东莞市儿童医院新生儿科行超声引导下静脉导管插入术的68例新生儿。每日应用POCUS监测导管尖端位置,直至导管拔除。导管尖端移位范围超过下腔静脉与右心房交点±0.5 cm被视为位置不当。

结果

64例新生儿导管尖端发生移位(94.1%,64/68),移位距离中位数为0.4 cm(最小值-0.2 cm,最大值1.2 cm)。10例新生儿因移位导致位置不当(14.7%,10/68)。移位通常发生在置管后2 - 4天,第2天、第3天和第4天的移位率分别为94.1%(64/68)、90.6%(58/64)和98.3%(59/60),置管后第9天仍可能发生。此外,位置不当主要发生在置管后第二天。监测过程中,58例新生儿导管尖端移位≥2次,导致252次移位和22次位置不当事件。其中,导管尖端向外从下腔静脉移位7次,均及时拔除。超声定位486次,间接避免X线检查486次。

结论

脐静脉置管后导管尖端易发生移位和位置不当,最常见于第2 - 4天。建议在脐静脉置管期间直至导管拔除,每日用POCUS监测尖端位置。