Applied Physics Department (Optometry Area) (V.N.-P., N.N.-P., M.J.G., E.Y.-P., H.P.-V.), Facultade de Óptica e Optometría, Universidade de Santiago de Compostela; and Optometry Group (M.J.G., E.Y.-P., H.P.-V.), Instituto de Investigación Sanitaria Santiago de Compostela (IDIS), Santiago de Compostela, Spain.
Eye Contact Lens. 2023 Nov 1;49(11):464-470. doi: 10.1097/ICL.0000000000001030. Epub 2023 Sep 15.
To compare the values of central corneal thickness, anterior chamber depth (ACD), lens thickness (LT), vitreous length (VL), axial length (AL), and corneal-compensated intraocular pressure performed with and without contact lenses (CL).
Forty volunteer participants (16 men and 24 women, 24.2±2.9 years) were recruited. In a single visit, participants underwent autorefraction, keratometry, biometry, topography, pachymetry, and tonometry with the naked eye (without CL). Then, biometry, pachymetry, and tonometry were repeated twice wearing two CL (somofilcon A and nesofilcon A) fitted in a random order.
Many of the ocular biometric values were affected by the use of CL during measurements (paired t test; all P ≤0.003), except for LT and VL (Wilcoxon test; both P ≥0.272). Corneal-compensated intraocular pressure was also affected by contact lens wear during measurements, obtaining lower values when wearing them (Wilcoxon test; all P ≤0.001).
Central corneal thickness, ACD, AL, and corneal-compensated intraocular pressure measurements cannot be performed while wearing CL. However, LT and VL measurements were not affected by any contact lens use. In addition, it was observed that ACD results from both devices are not interchangeable either when measured with the naked eye or using any contact lens.
比较角膜中央厚度、前房深度(ACD)、晶状体厚度(LT)、玻璃体长度(VL)、眼轴长度(AL)和角膜补偿眼压在使用和不使用接触镜(CL)时的测量值。
招募了 40 名志愿者参与者(16 名男性和 24 名女性,24.2±2.9 岁)。在单次就诊中,参与者接受了自动折射、角膜曲率计、生物测量、地形图、角膜测厚和眼压计检查(不戴 CL)。然后,两次佩戴两种 CL(硅水凝胶 A 和水凝胶 A),随机顺序重复进行生物测量、角膜测厚和眼压计检查。
在测量过程中使用 CL 会影响许多眼部生物测量值(配对 t 检验;所有 P≤0.003),但 LT 和 VL 除外(Wilcoxon 检验;两者 P≥0.272)。在测量过程中使用接触镜也会影响角膜补偿眼压,佩戴接触镜时获得的眼压值较低(Wilcoxon 检验;所有 P≤0.001)。
不能在佩戴 CL 的情况下进行中央角膜厚度、ACD、AL 和角膜补偿眼压的测量。然而,LT 和 VL 的测量值不受任何接触镜使用的影响。此外,观察到两种设备的 ACD 结果在裸眼或使用任何接触镜测量时都不能互换。