Ledoux Léa, Zirem Yanis, Renaud Florence, Duponchel Ludovic, Salzet Michel, Ogrinc Nina, Fournier Isabelle
Univ. Lille, Inserm, CHU Lille, U1192 - Protéomique Réponse Inflammatoire Spectrométrie de Masse - PRISM, F-59000 Lille, France.
Unité mixte SIRIC CURAMUS et 938, Inserm. Université Paris Sorbonne, France.
Analyst. 2023 Oct 5;148(20):4982-4986. doi: 10.1039/d3an01096a.
In this study, we conducted a direct comparison of water-assisted laser desorption ionization (WALDI) and matrix-assisted laser desorption ionization (MALDI) mass spectrometry imaging, with MALDI serving as the benchmark for label-free molecular tissue analysis in biomedical research. Specifically, we investigated the lipidomic profiles of several biological samples and calculated the similarity of detected peaks and Pearson's correlation of spectral profile intensities between the two techniques. We show that, overall, MALDI MS and WALDI MS present very close lipidomic analyses and that the highest similarity is obtained for the norharmane MALDI matrix. Indeed, for norharmane in negative ion mode, the lipidomic spectra revealed 100% similarity of detected peaks and over 0.90 intensity correlation between both technologies for five samples. The MALDI-MSI positive ion lipid spectra displayed more than 83% similarity of detected peaks compared to those of WALDI-MSI. However, we observed a lower percentage (77%) of detected peaks when comparing WALDI-MSI with MALDI-MSI due to the rich WALDI-MSI lipid spectra. Despite this difference, the global lipidomic spectra showed high consistency between the two technologies, indicating that they are governed by similar processes. Thanks to this similarity, we can increase datasets by including data from both modalities to either co-train classification models or obtain cross-interrogation.
在本研究中,我们对水辅助激光解吸电离(WALDI)和基质辅助激光解吸电离(MALDI)质谱成像进行了直接比较,其中MALDI作为生物医学研究中无标记分子组织分析的基准。具体而言,我们研究了几种生物样品的脂质组学谱,并计算了两种技术之间检测到的峰的相似度以及光谱轮廓强度的皮尔逊相关性。我们表明,总体而言,MALDI MS和WALDI MS呈现出非常相近的脂质组学分析结果,并且对于去氢骆驼蓬碱MALDI基质获得了最高的相似度。实际上,对于负离子模式下的去氢骆驼蓬碱,脂质组学光谱显示,两种技术对五个样品检测到的峰相似度为100%,强度相关性超过0.90。与WALDI-MSI相比,MALDI-MSI正离子脂质光谱检测到的峰相似度超过83%。然而,由于WALDI-MSI脂质光谱丰富,将WALDI-MSI与MALDI-MSI进行比较时,我们观察到检测到的峰的百分比更低(77%)。尽管存在这种差异,但整体脂质组学光谱显示这两种技术之间具有高度一致性,表明它们受相似过程的支配。由于这种相似性,我们可以通过纳入来自两种模式的数据来增加数据集,以共同训练分类模型或进行交叉询问。