无镜式增强现实定位在神经外科中的应用。
Application of Glasses-Free Augmented Reality Localization in Neurosurgery.
机构信息
Department of Neurosurgery, Hangzhou Xixi Hospital Affiliated to Zhejiang University School of Medicine, Hangzhou, China.
Department of Radiology, Hangzhou Xixi Hospital Affiliated to Zhejiang University School of Medicine, Hangzhou, China.
出版信息
World Neurosurg. 2023 Dec;180:e296-e301. doi: 10.1016/j.wneu.2023.09.064. Epub 2023 Sep 25.
OBJECTIVE
The accurate localization of intracranial lesions is critical in neurosurgery. Most surgeons locate the vast majority of neurosurgical sites through skull surface markers, combined with neuroimaging examination and marking lines. This project's primary purpose was to develop an augmented reality (AR) technology or tool that can be used for surgical positioning using the naked eye.
METHODS
Brain models were predesigned with intracranial lesions using computerized tomography scan, and Digital Imaging and Communications in Medicine data were segmented and modeled by 3D slicer software. The processed data were imported into a smartphone 3D viewing software application (Persp 3D) and were used by a Remebot surgical robot. The localization of intracranial lesions was performed, and the AR localization error was calculated compared with standard robot localization.
RESULTS
After mastering the AR localization registration method, surgeons achieved an average localization error of 1.39 ± 0.82 mm.
CONCLUSIONS
The error of AR positioning technology in surgical simulation tests based on brain modeling was millimeter level, which has verified the feasibility of clinical application. More efficient registration remains a need that should be addressed.
目的
在神经外科中,准确定位颅内病变至关重要。大多数外科医生通过颅骨表面标记物,结合神经影像学检查和标记线来定位大多数神经外科部位。本项目的主要目的是开发一种增强现实(AR)技术或工具,可用于通过肉眼进行手术定位。
方法
使用计算机断层扫描预先设计带有颅内病变的脑模型,并使用 3D slicer 软件对数字成像和通信中的医学数据进行分割和建模。处理后的数据被导入智能手机 3D 查看软件应用程序(Persp 3D),并由 Remebot 手术机器人使用。对颅内病变进行定位,并计算 AR 定位与标准机器人定位的误差。
结果
掌握 AR 定位注册方法后,外科医生的平均定位误差为 1.39±0.82mm。
结论
基于脑建模的手术模拟测试中,AR 定位技术的误差为毫米级,这验证了其临床应用的可行性。仍需要更高效的注册。