Smartphone-Based LiDAR Application for Easy and Accurate Wound Size Measurement.

作者信息

Song Bokeun, Kim Jeonghee, Kwon Hyeokjae, Kim Sunje, Oh Sang-Ha, Ha Yooseok, Song Seung-Han

机构信息

Department of Plastic and Reconstructive Surgery, Chungnam National University Hospital, Daejeon 35015, Republic of Korea.

Department of Medical Science, College of Medicine, Chungnam National University, Daejeon 35015, Republic of Korea.

出版信息

J Clin Med. 2023 Sep 19;12(18):6042. doi: 10.3390/jcm12186042.

Abstract

The accurate assessment of wound size is a critical step in advanced wound care management. This study aims to introduce and validate a Light Detection and Ranging (LiDAR) technique for measuring wound size. Twenty-eight wounds treated from December 2022 to April 2023 at the Chungnam National University Hospital were analyzed. All the wounds were measured using three techniques: conventional ruler methods, the LiDAR technique, and ImageJ analysis. Correlation analysis, linear regression, and Bland-Altman plot analysis were performed to validate the accuracy of the novel method. The measurement results (mean ± standard deviation) obtained using the ruler method, LiDAR technique, and ImageJ analysis were 112.99 ± 110.07 cm, 73.59 ± 72.97 cm, and 74.29 ± 72.15 cm, respectively. The Pearson correlation coefficient was higher for the LiDAR application (0.995) than for the conventional ruler methods (mean difference, -5.0000 cm), as was the degree of agreement (mean difference, 38.6933 cm). Wound size measurement using LiDAR is a simple and reliable method that will enable practitioners to conveniently assess wounds with a flattened and irregular shape with higher accuracy. However, non-flattened wounds cannot be assessed owing to the technical limitations of LiDAR.

摘要

准确评估伤口大小是高级伤口护理管理中的关键步骤。本研究旨在介绍并验证一种用于测量伤口大小的光探测与测距(LiDAR)技术。对2022年12月至2023年4月在忠南国立大学医院治疗的28处伤口进行了分析。所有伤口均使用三种技术进行测量:传统直尺法、LiDAR技术和ImageJ分析。进行了相关性分析、线性回归分析和Bland-Altman图分析,以验证该新方法的准确性。使用直尺法、LiDAR技术和ImageJ分析获得的测量结果(平均值±标准差)分别为112.99±110.07平方厘米、73.59±72.97平方厘米和74.29±72.15平方厘米。LiDAR应用的Pearson相关系数(0.995)高于传统直尺法(平均差值为-5.0000平方厘米),一致性程度(平均差值为38.6933平方厘米)也是如此。使用LiDAR测量伤口大小是一种简单可靠的方法,能够让从业者更方便地以更高的准确性评估扁平且形状不规则的伤口。然而,由于LiDAR的技术限制,无法评估未扁平化的伤口。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4867/10531604/910813b000bd/jcm-12-06042-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验