Is a smartphone application as accurate as a traditional goniometer for assessing finger joint angles in Dupuytren's disease?

作者信息

O'Brien Henry, Paniker Jay, Brown Liam

机构信息

Rotherham Hospital, Moorgate Road, Rotherham, S60 2UD, United Kingdom.

Rotherham Hospital, Moorgate Road, Rotherham, S60 2UD, United Kingdom.

出版信息

Hand Surg Rehabil. 2023 Dec;42(6):536-540. doi: 10.1016/j.hansur.2023.09.003. Epub 2023 Sep 29.

Abstract

OBJECTIVES

Smartphone applications are available using the smartphone accelerometer to measure angles. This study compared finger joint angle measurements using a traditional goniometer (TG) versus a smartphone application (SPA) in a group of patients with Dupuytren's disease, to determine if they were equally accurate. This was a clinical measurement case series. A group of patients with Dupuytren's disease were invited to participate.

MATERIAL AND METHODS

The Apple iPhone 7 "Measure" application was used. Participants were asked to place their hand as flat as possible on the table in a clinic room. Each joint of the ring and little fingers was measured using the TG and the SPA. Power calculation by paired t-test determined a sample size of 25.

RESULTS

28 hands were measured. The Pearson correlation coefficients for the various finger joints were: metacarpophalangeal joint (MCPJ), 0.989; proximal interphalangeal joint (PIPJ), 0.997; and distal interphalangeal joint (DIPJ), 0.977. The root mean square errors (RMSE) for the various joints were: MCPJ, 3.190; PIPJ, 2.019; and DIPJ, 1.897. Bland-Altman analysis showed the two methods to be in agreement.

CONCLUSION

The SPA was consistent, reliable and in agreement with a TG for assessing ring and little finger joint angles. The SPA could be used by healthcare professionals instead of a TG.