Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, 518055, China.
School of Physics and Electronics-Electrical Engineering, Xiangnan University, Chenzhou, 423000, China.
Eur J Nucl Med Mol Imaging. 2024 Jan;51(2):346-357. doi: 10.1007/s00259-023-06458-z. Epub 2023 Oct 2.
Positron emission tomography/magnetic resonance imaging (PET/MRI) is a powerful tool for brain imaging, but the spatial resolution of the PET scanners currently used for brain imaging can be further improved to enhance the quantitative accuracy of brain PET imaging. The purpose of this study is to develop an MR-compatible brain PET scanner that can simultaneously achieve a uniform high spatial resolution and high sensitivity by using dual-ended readout depth encoding detectors.
The MR-compatible brain PET scanner, named SIAT bPET, consists of 224 dual-ended readout detectors. Each detector contains a 26 × 26 lutetium yttrium oxyorthosilicate (LYSO) crystal array of 1.4 × 1.4 × 20 mm crystal size read out by two 10 × 10 silicon photomultiplier (SiPM) arrays from both ends. The scanner has a detector ring diameter of 376.8 mm and an axial field of view (FOV) of 329 mm. The performance of the scanner including spatial resolution, sensitivity, count rate, scatter fraction, and image quality was measured. Imaging studies of phantoms and the brain of a volunteer were performed. The mutual interferences of the PET insert and the uMR790 3 T MRI scanner were measured, and simultaneous PET/MRI imaging of the brain of a volunteer was performed.
A spatial resolution of better than 1.5 mm with an average of 1.2 mm within the whole FOV was obtained. A sensitivity of 11.0% was achieved at the center FOV for an energy window of 350-750 keV. Except for the dedicated RF coil, which caused a ~ 30% reduction of the sensitivity of the PET scanner, the MRI sequences running had a negligible effect on the performance of the PET scanner. The reduction of the SNR and homogeneity of the MRI images was less than 2% as the PET scanner was inserted to the MRI scanner and powered-on. High quality PET and MRI images of a human brain were obtained from simultaneous PET/MRI scans.
The SIAT bPET scanner achieved a spatial resolution and sensitivity better than all MR-compatible brain PET scanners developed up to date. It can be used either as a standalone brain PET scanner or a PET insert placed inside a commercial whole-body MRI scanner to perform simultaneous PET/MRI imaging.
正电子发射断层扫描/磁共振成像(PET/MRI)是一种强大的脑成像工具,但目前用于脑成像的 PET 扫描仪的空间分辨率可以进一步提高,以提高脑 PET 成像的定量准确性。本研究的目的是开发一种兼容磁共振的脑 PET 扫描仪,该扫描仪可以通过使用双端读出深度编码探测器同时实现均匀的高空间分辨率和高灵敏度。
这种名为 SIAT bPET 的兼容磁共振的脑 PET 扫描仪由 224 个双端读出探测器组成。每个探测器包含一个 26×26 的硅酸镥(Lutetium Yttrium Oxyorthosilicate,LYSO)晶体阵列,晶体尺寸为 1.4×1.4×20mm,由两个 10×10 的硅光电倍增管(Silicon Photomultiplier,SiPM)阵列从两端读取。扫描仪的探测器环直径为 376.8mm,轴向视野(Field of View,FOV)为 329mm。测量了扫描仪的性能,包括空间分辨率、灵敏度、计数率、散射分数和图像质量。对 phantom 和志愿者大脑进行了成像研究。测量了 PET 插入件与 uMR790 3T MRI 扫描仪之间的相互干扰,并对志愿者大脑进行了同时 PET/MRI 成像。
在整个 FOV 内获得了优于 1.5mm、平均为 1.2mm 的空间分辨率。在中心 FOV 内,对于 350-750keV 的能窗,灵敏度达到 11.0%。除了专用的射频线圈导致 PET 扫描仪灵敏度降低约 30%外,运行的 MRI 序列对 PET 扫描仪的性能几乎没有影响。当 PET 扫描仪插入 MRI 扫描仪并开启电源时,MRI 图像的 SNR 和均匀性降低小于 2%。从同时的 PET/MRI 扫描中获得了高质量的人脑 PET 和 MRI 图像。
SIAT bPET 扫描仪实现了优于迄今为止开发的所有兼容磁共振的脑 PET 扫描仪的空间分辨率和灵敏度。它既可以作为独立的脑 PET 扫描仪使用,也可以作为放置在商用全身 MRI 扫描仪内部的 PET 插入件,以进行同时的 PET/MRI 成像。