A simple device for the direct measurement of mean arterial pressure and for calibration of arterial pressure monitors.

作者信息

Hickey D D

出版信息

J Med Eng Technol. 1986 Jul-Aug;10(4):188-92. doi: 10.3109/03091908609022907.

Abstract

A simple, inexpensive mechanical device is described for the direct continuous measurement of mean arterial pressure. It can be assembled in minutes using a few dollars worth of sterile components commonly available in an intensive care unit. Mean arterial pressure determined in patients with this device differs by 1.6 mmHg +/- 1.1 SD from mean arterial pressure simultaneously measured by properly calibrated Hewlett-Packard electronic monitors (N = 5), a non-significant difference. Similar agreement was obtained when the device was tested using a commercial blood pressure system calibrator as the pressure source. This device also allows rapid static calibration of electronic pressure transducers across the full range of human blood pressure using a 3 ml syringe as the pressure source and an aneroid gauge, or mercury column, as the pressure standard.

摘要

本文介绍了一种简单、廉价的机械装置,用于直接连续测量平均动脉压。使用重症监护病房中常见的价值几美元的无菌组件,几分钟内即可组装完成。使用该装置测定的患者平均动脉压与经适当校准的惠普电子监护仪同时测量的平均动脉压相差1.6 mmHg±1.1标准差(N = 5),差异无统计学意义。当使用商用血压系统校准器作为压力源对该装置进行测试时,也得到了类似的一致性结果。该装置还允许使用3 ml注射器作为压力源,以无液压力计或汞柱作为压力标准,在全人类血压范围内对电子压力传感器进行快速静态校准。