Am J Vet Res. 2023 Oct 23;85(1). doi: 10.2460/ajvr.23.05.0107. Print 2024 Jan 1.
To evaluate the agreement between the Tafonius large animal ventilator-integrated volumetric capnography (vCap) software and the Respironics NICO noninvasive cardiac output monitor reference system.
Data were collected from 56 healthy adult horses undergoing general anesthesia.
Animals were placed under general anesthesia and connected to the Tafonius large animal ventilator circle system. A flow partitioning device with CO2 and flow sensors was utilized to couple the endotracheal tube to the NICO monitor. Tafonius CO2 and flow sensors are incorporated into the Y-piece of the breathing circuit. Arterial blood samples were collected to determine the partial pressure of arterial carbon dioxide (PaCO2) immediately before data collection. The PaCO2 was input into the Tafonius and NICO monitor, and dead space ventilation (%Vd), end-tidal CO2 partial pressure (ETco2), mixed-expired CO2 partial pressure (Peco2), and expired tidal volume (Vt) were calculated over a single breath. Multiple measurements were completed for each patient, with a total of 200 paired data points collected for analysis. Data were assessed for normality, and Bland-Altman analysis was performed. Bias and 95% limits of agreement were calculated.
The limits of agreement for %Vd of the ventilator-derived measurements fell within ± 10% of the NICO monitor reference method.
Our results indicate that, when compared to the NICO monitor method, the Tafonius-integrated vCap software provides clinically acceptable values of Peco2, Vt, and %Vd in healthy adult horses.
评估 Tafonius 大型动物呼吸机集成容积二氧化碳描记法(vCap)软件与 Respironics NICO 无创心输出量监测参考系统之间的一致性。
数据来自 56 头接受全身麻醉的健康成年马。
将动物置于全身麻醉下,并连接到 Tafonius 大型动物呼吸机环系统。使用带有二氧化碳和流量传感器的流量分配装置将气管内导管与 NICO 监测器耦合。Tafonius CO2 和流量传感器被整合到呼吸回路的 Y 型件中。在收集数据之前,采集动脉血样以确定动脉血二氧化碳分压(PaCO2)。将 PaCO2 输入 Tafonius 和 NICO 监测器,计算单次呼吸时的通气死腔百分比(%Vd)、呼气末二氧化碳分压(ETco2)、混合呼出二氧化碳分压(Peco2)和呼出潮气量(Vt)。对每个患者完成多次测量,共收集 200 对数据点进行分析。评估数据的正态性,并进行 Bland-Altman 分析。计算偏差和 95%一致性界限。
呼吸机衍生测量的 %Vd 的一致性界限在 NICO 监测器参考方法的±10%范围内。
我们的结果表明,与 NICO 监测器方法相比,Tafonius 集成 vCap 软件在健康成年马中提供了可接受的 Peco2、Vt 和 %Vd 值。