Department of Cardiac Anaesthesiology, Sri Jayadeva Institute of Cardiovascular Sciences and Research, Jayanagar, Bengaluru, Karnataka, India.
Ann Card Anaesth. 2023 Oct-Dec;26(4):380-385. doi: 10.4103/aca.aca_9_23.
The present study was conducted to validate cardiac output (CO) and cardiac index (CI) obtained from electrical cardiometry (EC) ICON with transthoracic echocardiography (TTE) in postoperative pediatric cardiac surgical patients.
A prospective observational study was conducted in 25 pediatric patients with age < 10 years who underwent elective cardiac surgery.
BlandAltman plot was constructed for interchangeability and Polar plot was constructed to know trending ability.
A total of 250 datasets were analyzed. Spearman's correlation coefficient for CO between ICON and TTE showed good positive correlation (r = 0.850, 95% confidence interval 0.81 to 0.881, P <.0001). Moderate positive correlation was observed between ICON and TTE for CI (r = 0.60, 95% confidence interval 0.515 to 0.674, P <.0001). Linear regression equations for CO and CI between ICON and TTE were: y = 0.5230 + 0.8078 X (R = 0.6597, P <.001) and y = 1.8350 + 0.5869 X (R = 0.3985, P <.001) [y- ICON ; X - TTE], respectively. BlandAltman plot for CO between ICON and TTE showed a bias of 0.3012 with limits of agreement (LOA) being -0.69 to 1.3 and for CI bias was 0.6939 with LOA-2.1 to 3.5. Polar plot analysis showed an angular bias of 8.1750, with radial LOA being -13.74° to 30.08° for CO and angular bias of 6.6931, with radial LOA being -15.69° to 29.07° for CI.
ICON monitor-derived parameters are not interchangeable with the values derived from TTE. However, the ICON monitor demonstrated a good trending ability for both CO and CI.
本研究旨在验证经胸超声心动图(TTE)与心电机械图(EC) ICON 在小儿心脏手术后患者中测量心输出量(CO)和心指数(CI)的准确性。
本前瞻性观察研究纳入了 25 名年龄<10 岁的择期心脏手术患儿。
采用 Bland-Altman 图进行可互换性分析,采用极坐标图分析趋势能力。
共分析了 250 组数据集。ICON 与 TTE 测量 CO 的 Spearman 相关系数显示出良好的正相关性(r = 0.850,95%置信区间为 0.81 至 0.881,P<.0001)。ICON 与 TTE 测量 CI 时,中度正相关(r = 0.60,95%置信区间为 0.515 至 0.674,P<.0001)。ICON 与 TTE 测量 CO 和 CI 的线性回归方程分别为:y = 0.5230 + 0.8078 X(R = 0.6597,P<.001)和 y = 1.8350 + 0.5869 X(R = 0.3985,P<.001)[y- ICON ;X - TTE]。ICON 与 TTE 测量 CO 的 Bland-Altman 图显示出 0.3012 的偏倚,一致性界限(LOA)为-0.69 至 1.3,CI 的偏倚为 0.6939,LOA 为-2.1 至 3.5。极坐标图分析显示 CO 的角度偏倚为 8.1750,径向 LOA 为-13.74°至 30.08°,CI 的角度偏倚为 6.6931,径向 LOA 为-15.69°至 29.07°。
ICON 监测仪衍生参数与 TTE 衍生值不可互换。然而,ICON 监测仪在 CO 和 CI 方面均显示出良好的趋势能力。